Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs Ionizing Radiation (3)
IONIZING RADIATION
Absorbed dose rate to tissue in 0.07 mm depth (3)
GEOGRAPHIC LOCATION
APMP (3)
Japan (3)
Result of the search
Your query 'Japan' produced 3 results New search

 
Japan, NMIJ AIST (National Metrology Institute of Japan)
Complete CMCs in Ionizing Radiation for Japan (.PDF file)
Absorbed dose rate to tissue in 0.07 mm depth. Ionization chamber, 1.1E-05 Gy s-1 to 1.1E-05 Gy s-1
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.8
Irradiation in a calibrated field
Beta radiation: Sr-90/Y-90, 460 MBq (2006.2), ISO 6980
Reference standard: Primary extrapolation chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1016
Absorbed dose rate to tissue in 0.07 mm depth. Ionization chamber, 3.5E-05 Gy s-1 to 3.5E-05 Gy s-1
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.8
Irradiation in a calibrated field
Beta radiation: Kr-85, 3.7 GBq (2006.2), ISO 6980
Reference standard: Primary extrapolation chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1017
Absorbed dose rate to tissue in 0.07 mm depth. Ionization chamber, 2.0E-06 Gy s-1 to 2.0E-06 Gy s-1
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 4.8
Irradiation in a calibrated field
Beta radiation: Pm-147, 3.7 GBq (2006.2), ISO 6980
Reference standard: Primary extrapolation chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1018