Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (8)
PHYSICS
Materials
Photometry
Properties of detectors and sources (8)
Responsivity, spectral, power (8)
CHEMICAL MATERIAL
monoelemental solution
GEOGRAPHIC LOCATION
EURAMET (8)
United Kingdom (8)
Result of the search
Your query 'United Kingdom' produced 8 results New search

 
United Kingdom, NPL (National Physical Laboratory)
Complete CMCs in Photometry and Radiometry for United Kingdom (.PDF file)
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.6 to 3.2, depending on wavelength
Double grating monochromator
Wavelength: 200 nm to 210 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.0
Double grating monochromator
Wavelength: 211 nm to 239 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.7 at 240 nm, 0.5 above 240 nm
Double grating monochromator
Wavelength: 240 nm to 315 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.3
Double grating monochromator
Wavelength: 316 nm to 404 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.1
Double grating monochromator
Wavelength: 405 nm to 919 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.3
Double grating monochromator
Wavelength: 920 nm to 1000 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.3
Double grating monochromator
Wavelength: 1001 nm to 1400 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.3 to 0.4 (depending on wavelength)
Double grating monochromator
Wavelength: 1401 nm to 1800 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017