United Kingdom, NPL (National Physical Laboratory)
Complete CMCs in Ionizing Radiation for United Kingdom (.PDF file)
Air kerma. Therapy level secondary standard dosemeter, 1.00E-01 Gy to 5.00E+00 Gy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.2
Calibration against primary standard
X-rays: X-rays ([0.024 to 1] mm Al HVL, [8.5 to 50] kV)
Reference standard: Free air chamber
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1021
Air kerma. Therapy level secondary standard dosemeter, 1.00E-01 Gy to 5.00E+00 Gy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.2
Calibration against primary standard
X-rays: X-rays ([1 to 20] mm Al HVL, [50 to 280] kV)
Reference standard: Free air chamber
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1022
Air kerma. Therapy level secondary standard dosemeter, 1.00E-01 Gy to 5.00E+00 Gy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.4
Calibration against primary standard
Co-60: Co-60 gamma rays
Reference standard: Cavity standard
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1023
Air kerma. Protection level secondary standard dosemeter, 5.00E-03 mGy to 2.00E+02 mGy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.6
Calibration against transfer standard
X-rays: X-rays (ISO 4037 narrow series) [10 to 30] kV, [0.058 to 1.21] mm Al HVL
Reference standard: Free air chamber
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1026
Air kerma. Protection level secondary standard dosemeter, 5.00E-03 mGy to 2.00E+02 mGy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.6
Calibration against transfer standard
X-rays: X-rays (ISO 4037 narrow series) [40 to 250] kV, [0.085 to 6.26] mm Cu HVL
Reference standard: Free air chamber
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1027
Air kerma. Protection level secondary standard dosemeter, 5.00E-03 mGy to 2.00E+02 mGy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.7
Calibration against transfer standard
Am-241: Am-241 gamma rays
Reference standard: Free air chamber
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1028
Air kerma. Protection level secondary standard dosemeter, 5.00E-03 mGy to 2.00E+02 mGy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.4
Calibration against transfer standard
Cs-137: Cs-137 gamma rays
Reference standard: Cavity standard
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1029
Air kerma. Protection level secondary standard dosemeter, 5.00E-03 mGy to 2.00E+02 mGy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.4
Calibration against transfer standard
Co-60: Co-60 gamma rays
Reference standard: Cavity standard
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1030
Air kerma. Mammographic secondary standard dosemeter, 1.00E-02 Gy to 5.00E+00 Gy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.1
Calibration against primary standard
X-rays: X-rays (IEC 61267), molybdenum anode, 28 kV
Reference standard: Free air chamber
Source of traceability: NPL
RQR-M2 (28 kV, no filter). RQB-M2 (28 kV + 45 mm breast equivalent phantom). Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1036
Air kerma. Diagnostic secondary standard dosemeter, 1.00E-02 Gy to 5.00E+00 Gy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.6
Calibration against primary standard
X-rays: X-rays (IEC 61267) tungsten anode, [40 to 150] kV
Reference standard: Free air chamber
Source of traceability: NPL
Diagnostic entrance qualities: RQR2, 3, 4, 5, 6, 7, 8, 9, 10 . Diagnostic exit qualities: RQA2, 3, 4, 5, 6, 7, 8, 9, 10. Fluoroscopy qualities: RQC3, 5, 8. CT qualities: RQT8,9,10. Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1037