Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs Ionizing Radiation (15)
PHYSICS
Materials
Photometry
Impedance up to the MHz range
IONIZING RADIATION
Absorbed dose to water (15)
CHEMICAL MATERIAL
monoelemental solution
GEOGRAPHIC LOCATION
EURAMET (15)
United Kingdom (8)
Denmark (7)
Result of the search
Your query 'United Kingdom' produced 15 results New search

 
Denmark, DTU (Technical University of Denmark)
Complete CMCs in Ionizing Radiation for Denmark (.PDF file)
Absorbed dose to water. Passive dosimeter (Alanine, Radiochromic film, Perspex), 0.2 kGy to 100 kGy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.9
Irradiation in calibrated photon field
Photons: Irradiation in reference Co-60 gamma ray field (Gamma cell)
Reference standard: Photon beam graphite calorimeter at National Physical Laboratory
Source of traceability: National Physical Laboratory, United Kingdom
Approved on 20 April 2015
Internal NMI service identifier: EUR-RAD-DTU-1001
Absorbed dose to water. Alanine pellet / EPR dosimeter, 0.2 kGy to 100 kGy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.2
Irradiation of transfer dosimeters in phantoms
Photons: Irradiation in customer's Co-60 gamma field
Reference standard: Calibrated Co-60 gamma field
Source of traceability: National Physical Laboratory, United Kingdom
Approved on 20 April 2015
Internal NMI service identifier: EUR-RAD-DTU-1002
Absorbed dose to water. Graphite calorimeter, 1.5 kGy to 15 kGy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 4.1
Irradiation in calibrated electron field
Electrons: Irradiation in customer's electron field (6 to 10 MeV)
Reference standard: Calibrated Co-60 gamma field
Source of traceability: National Physical Laboratory, United Kingdom
Approved on 20 April 2015
Internal NMI service identifier: EUR-RAD-DTU-1003
Absorbed dose to water. Polystyrene calorimeter, 3 kGy to 10 kGy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 4.0
Irradiation in calibrated electron field
Electrons: Irradiation in customer's electron field (1.5 to 10 MeV)
Reference standard: Calibrated Co-60 gamma field
Source of traceability: National Physical Laboratory, United Kingdom
Approved on 20 April 2015
Internal NMI service identifier: EUR-RAD-DTU-1004
Absorbed dose to water. Alanine pellet / EPR dosimeter, 0.2 kGy to 100 kGy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.5
Irradiation of transfer dosimeters in phantoms
Electrons: Irradiation in customer's electron field (1.5 to 10 MeV)
Reference standard: Calibrated Co-60 gamma field
Source of traceability: National Physical Laboratory, United Kingdom
Approved on 20 April 2015
Internal NMI service identifier: EUR-RAD-DTU-1007
Absorbed dose to water. Alanine film / EPR dosimeter, 1 kGy to 120 kGy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.6
Irradiation of transfer dosimeters in phantoms
Electrons: Irradiation in customer's electron field (0.1 to 10 MeV)
Reference standard: Calibrated Co-60 gamma field
Source of traceability: National Physical Laboratory, United Kingdom
Approved on 20 April 2015
Internal NMI service identifier: EUR-RAD-DTU-1008
Absorbed dose to water. Risø B3 film dosimeter, 10 kGy to 100 kGy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 8.2
Irradiation in calibrated electron field
Electrons: Irradiation in customer's electron field (0.1 to 10 MeV)
Reference standard: Calibrated Co-60 gamma field
Source of traceability: National Physical Laboratory, United Kingdom
Approved on 20 April 2015
Internal NMI service identifier: EUR-RAD-DTU-1009
 
United Kingdom, NPL (National Physical Laboratory)
Complete CMCs in Ionizing Radiation for United Kingdom (.PDF file)
Absorbed dose to water. Therapy Level Secondary Standard Dosemeter, 1.00E-01 Gy to 5.00E+00 Gy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.5
Calibration against working standard in water phantom
Electrons: [4 to 20] MeV (R50water = [1.6 to 8.1] cm)
Reference standard: Electron Beam Graphite Calorimeter
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1009
Absorbed dose to water. Passive dosemeter, 5.00E+00 Gy to 1.00E+05 Gy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.2
Irradiation in calibrated photon field
Co-60: Co-60 gamma rays
Reference standard: Photon Beam Graphite Calorimeter
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1010
Absorbed dose to water. Alanine dosimetry, 5.00E+00 Gy to 1.00E+05 Gy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.8 to 2.0
Irradiation in photon field
Cs-137: Cs-137 gamma rays
Reference standard: Photon Beam Graphite Calorimeter
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1011
Absorbed dose to water. Alanine dosimetry, 5.00E+00 Gy to 1.00E+05 Gy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.8 to 2.0
Irradiation in photon field
Co-60: Co-60 gamma rays
Reference standard: Photon Beam Graphite Calorimeter
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1012
Absorbed dose to water. Alanine dosimetry, 5.00E+00 Gy to 1.00E+05 Gy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.8 to 2.0
Irradiation in photon field
Photons, high energy: Photons generated above 2 MeV
Reference standard: Photon Beam Graphite Calorimeter
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1013
Absorbed dose to water. Alanine dosimetry, 5.00E+00 Gy to 1.00E+05 Gy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.8 to 2.0
Irradiation in electron field
Electrons: Electrons, energy above 4 MeV
Reference standard: Photon Beam Graphite Calorimeter
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1014
Absorbed dose to water. Dichromate dosimetry, 2.00E+03 Gy to 5.5E+04 Gy
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.9 to 2.2
Irradiation in photon field
Co-60: Co-60 gamma rays
Reference standard: Photon Beam Graphite Calorimeter
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1019
Absorbed dose to water. Ophthalmic applicators, 2.00E-01 Gy/s to 1.00E+00 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 7
Calibration against alanine transfer standard
Sr-90 or Ru-106: Sr-90 or Ru-106 beta rays
Reference standard: Photon Beam Graphite Calorimeter
Source of traceability: NPL
Approved on 02 July 2015
Internal NMI service identifier: EUR-RAD-NPL-1041