Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (7)
PHYSICS
Materials
Photometry
Properties of detectors and sources (7)
Radiance, spectral (7)
CHEMICAL MATERIAL
monoelemental solution
GEOGRAPHIC LOCATION
EURAMET (7)
United Kingdom (7)
Result of the search
Your query 'United Kingdom' produced 7 results New search

 
United Kingdom, NPL (National Physical Laboratory)
Complete CMCs in Photometry and Radiometry for United Kingdom (.PDF file)
Radiance, spectral. Tungsten lamp, 1E-05 W/(m2 sr nm) to 0.1 W/(m2 sr nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.5 to 1.9 (with wavelength)
Spectroradiometer
Wavelength range: 250 nm to 300 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Radiance, spectral. Tungsten lamp, 1E-05 W/(m2 sr nm) to 0.1 W/(m2 sr nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.9
Spectroradiometer
Wavelength range: 300 nm to 315 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Radiance, spectral. Tungsten lamp, 1E-05 W/(m2 sr nm) to 0.1 W/(m2 sr nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.9 to 1.2 (with wavelength)
Spectroradiometer
Wavelength range: 315 nm to 430 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Radiance, spectral. Tungsten lamp, 1E-04 W/(m2 sr nm) to 1 W/(m2 sr nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.2 to 0.8, with wavelength
Spectroradiometer
Wavelength range: 430 nm to 600 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Radiance, spectral. Tungsten lamp, 1E-04 W/(m2 sr nm) to 1 W/(m2 sr nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.8
Spectroradiometer
Wavelength range: 600 nm to 1599 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Radiance, spectral. Tungsten lamp, 1E-04 W/(m2 sr nm) to 1 W/(m2 sr nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.8 to 2.0 (with wavelength)
Spectroradiometer
Wavelength range: 1600 nm to 2350 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Radiance, spectral. Tungsten lamp, 5E-04 W/(m2 sr nm) to 1 W/(m2 sr nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.4
Spectroradiometer
Wavelength range: 2351 nm to 2500 nm
Bandwidth: < 20 nm
Other types of source can also be measured