Japan, NMIJ AIST (National Metrology Institute of Japan)
Complete CMCs in Ionizing Radiation for Japan (.PDF file)
Air kerma rate. Ionization chamber, 2.5E-06 Gy/s to 1.0E-02 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.8
Calibration in air against primary standard
X ray, 10 kV to 50 kV: BIPM quality, 10 kV to 50 kV, 0.5 m to 3 m
Reference standard: Primary free air chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1001
Air kerma rate. Ionization chamber, 9.0E-09 Gy/s to 1.0E-02 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.8
Calibration in air against primary standard
X-ray, 10 kV to 50 kV: ISO 4037, Narrow Series, 10 kV to 50 kV, 0.5 m to 5 m
Reference standard: Primary free air chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1002
Air kerma rate. Ionization chamber, 4.0E-07 Gy/s to 2.0E-03 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.2
Calibration in air against primary standard
X ray, 50 kV to 420 kV: BIPM quality, 100 kV to 250 kV, 1.2 m to 5 m
Reference standard: Primary free air chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1003
Air kerma rate. Ionization chamber, 9.0E-09 Gy/s to 1.0E-05 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.2
Calibration in air against primary standard
X ray, 50 kV to 420 kV: ISO 4037, Narrow Spectrum, 50 kV to 300 kV, 1.2 m to 5 m
Reference standard: Primary free air chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1004
Air kerma rate. Ionization chamber, 9.0E-09 Gy/s to 1.0E-06 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.2
Calibration in air against primary standard
X ray, 50 kV to 420 kV: ISO 4037, Low kerma rate, 50 kV to 300 kV, 1.2 m to 5 m
Reference standard: Primary free air chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1005
Air kerma rate. Ionization chamber, 2.0E-07 Gy/s to 2.0E-03 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.2
Calibration in air against primary standard
X ray, 50 kV to 420 kV: ISO 4037, High kerma rate, 50 kV to 300 kV, 1.2 m to 5 m
Reference standard: Primary free air chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1006
Air kerma rate. Ionization chamber, 2.0E-08 Gy/s to 1.0E-04 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.2
Calibration in air against primary standard
X ray, 50 kV to 420 kV: ISO 4037, Wide spectrum, 50 kV to 300 kV, 1.2 m to 5 m
Reference standard: Primary free air chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1007
Air kerma rate. Ionization chamber, 4.3E-07 Gy/s to 7.9E-02 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.8
Irradiation in a calibrated field
Co-60: 148 TBq (1998), 0.3 m to 6 m, W-disk attenuator
Reference standard: Primary cavity chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1008
Air kerma rate. Ionization chamber, 2.0E-07 Gy/s to 6.4E-06 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.1
Irradiation in a calibrated field
Co-60: 185 GBq (1997), 1.1 m to 6 m
Reference standard: Primary cavity chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1009
Air kerma rate. Ionization chamber, 2.0E-08 Gy/s to 6.4E-07 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.2
Irradiation in a calibrated field
Co-60: 18.5 GBq (1997), 1.1 m to 6 m
Reference standard: Primary cavity chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1010
Air kerma rate. Ionization chamber, 3.6E-09 Gy/s to 1.1E-07 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.6
Irradiation in a calibrated field
Co-60: 3.7 GBq (1997), 1.1 m to 6 m
Reference standard: Primary cavity chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1011
Air kerma rate. Ionization chamber, 9.0E-06 Gy/s to 5.1E-04 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.6
Irradiation in a calibrated field
Cs-137: 34 TBq (1993), 1 m to 4 m, Fe-disk attenuator
Reference standard: Primary cavity chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1012
Air kerma rate. Ionization chamber, 9.0E-08 Gy/s to 2.9E-06 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.5
Irradiation in a calibrated field
Cs-137: 222 GBq (1980), 1.1 m to 6 m
Reference standard: Primary cavity chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1013
Air kerma rate. Ionization chamber, 5.4E-09 Gy/s to 1.7E-07 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.0
Irradiation in a calibrated field
Cs-137: 18.5 GBq (1980), 1.1 m to 6 m
Reference standard: Primary cavity chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1014
Air kerma rate. Ionization chamber, 6.7E-10 Gy/s to 2.2E-08 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.4
Irradiation in a calibrated field
Cs-137: 1.85 GBq (1980), 1.1 m to 6 m
Reference standard: Primary cavity chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1015
Air kerma rate. Ionization chamber, 5.0E-06 Gy/s to 5.0E-03 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.8
Calibration in air against primary standard
X-ray, 10 kV to 50 kV: Mammography
Anode material: Mo
Filter: Mo 0.030 mm, Mo 0.032 mm, Rh 0.025 mm
Reference standard: Primary free air chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1019
Air kerma rate. Ionization chamber, 3.0E-05 Gy/s to 3.0E-03 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.8
Calibration in air against primary standard
X-ray, 10 kV to 50 kV: Mammography
Anode material: Rh
Filter: Rh 0.025 mm
Reference standard: Primary free air chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015
Internal NMI service identifier: APM-RAD-NMIJ/AIST-1020
Air kerma rate. Ionization chamber, 5.0E-05 Gy/s to 5.0E-03 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.8
Calibration in air against primary standard
X-ray, 10 kV to 50 kV: Mammography
Anode material: W
Filter: Rh 0.05 mm, Ag 0.05 mm, Al 0.5 mm, Al 0.7 mm
Reference standard: Primary free air chamber
Source of traceability: NMIJ/AIST
Approved on 17 April 2015