Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs Ionizing Radiation (4)
IONIZING RADIATION
Personal dose equivalent rate (4)
GEOGRAPHIC LOCATION
APMP (4)
Korea, Republic of (4)
Result of the search
Your query 'kriss' produced 4 results New search

 
Korea, Republic of, KRISS (Korea Research Institute of Standards and Science)
Complete CMCs in Ionizing Radiation for Korea, Republic of (.PDF file)
Personal dose equivalent rate. Neutron personal dosemeter, 2.0E-04 Sv/h to 6.0E-04 Sv/h
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 10
Irradiation with standard neutron source
Bare Cf-252: ISO 8529-3 (at 1 m from the source and for an angle of incidence of 0 deg)
Reference standard: Calibrated source
Source of traceability: KRISS
Approved on 01 June 2011
Internal NMI service identifier: APM-RAD-KRISS-3008
Personal dose equivalent rate. Neutron personal dosemeter, 1.5E-04 Sv/h to 1.5E-04 Sv/h
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 10
Irradiation with standard neutron source
Am-241/Be-9: ISO 8529-3 (at 1 m from the source and for an angle of incidence of 0 deg)
Reference standard: Calibrated source
Source of traceability: KRISS
Approved on 01 June 2011
Internal NMI service identifier: APM-RAD-KRISS-3009
Personal dose equivalent rate. Neutron personal dosemeter, 6.0E-05 Sv/h to 1.8E-04 Sv/h
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 10
Irradiation with standard neutron source
Cf-252, D2O-moderated: ISO 8529-3 (at 1 m from the source and for an angle of incidence of 0 deg)
Reference standard: Calibrated source
Source of traceability: KRISS
Approved on 01 June 2011
Internal NMI service identifier: APM-RAD-KRISS-3010
Personal dose equivalent rate. Personal dosimeter, 5.0E+00 mSv/h to 1.3E+02 mSv/h
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 5
Irradiation in calibrated field using a phantom
Beta radiation: Sr-90/Y-90, ISO 6980, 1.85 GBq (1994), 74 MBq (1994)
Reference standard: Extrapolation chamber
Source of traceability: KRISS
Approved on 01 June 2011
Internal NMI service identifier: APM-RAD-KRISS-1014