Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs Ionizing Radiation (11)
IONIZING RADIATION >
Surface emission rate per unit area (11)
Tl-204 (1)
Pb-210 (1)
Am-241 (1)
Cl-36 (1)
Co-60 (1)
Sr-90 (1)
Cs-137 (1)
Pm-147 (1)
C-14 (1)
Ni-63 (1)
GEOGRAPHIC LOCATION
APMP (11)
Japan (11)
Result of the search
Your query 'Japan' produced 11 results New search

 
Japan, NMIJ AIST (National Metrology Institute of Japan)
Complete CMCs in Ionizing Radiation for Japan (.PDF file)
Surface emission rate per unit area. Large area source, 2.0E+00 s-1 cm-2 to 1.00E+03 s-1 cm-2
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3
2 proportional counter
C-14: area source between 20 cm2 and 500 cm2, E > 0.5 keV
Reference standard: -particle counting with 2 windowless gas flow counter
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2171
Surface emission rate per unit area. Large area source, 2.0E+00 s-1 cm-2 to 1.00E+03 s-1 cm-2
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
2 proportional counter
Cl-36: area source between 20 cm2 and 500 cm2, E > 0.5 keV
Reference standard: -particle counting with 2 windowless gas flow counter
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2172
Surface emission rate per unit area. Large area source, 2.0E+00 s-1 cm-2 to 1.00E+03 s-1 cm-2
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
2 proportional counter
Co-60: area source between 20 cm2 and 500 cm2, E > 0.5 keV
Reference standard: -particle counting with 2 windowless gas flow counter
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2173
Surface emission rate per unit area. Large area source, 2.0E+00 s-1 cm-2 to 1.00E+03 s-1 cm-2
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
2 proportional counter
Ni-63: area source between 20 cm2 and 500 cm2, E > 0.5 keV
Reference standard: -particle counting with 2 windowless gas flow counter
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2174
Surface emission rate per unit area. Large area source, 2.0E+00 s-1 cm-2 to 1.00E+03 s-1 cm-2
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
2 proportional counter
Sr-90/Y-90: area source between 20 cm2 and 500 cm2, E > 0.5 keV
Reference standard: -particle counting with 2 windowless gas flow counter
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2175
Surface emission rate per unit area. Large area source, 2.0E+00 s-1 cm-2 to 1.00E+03 s-1 cm-2
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
2 proportional counter
Tc-99: area source between 20 cm2 and 500 cm2, E > 0.5 keV
Reference standard: -particle counting with 2 windowless gas flow counter
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2176
Surface emission rate per unit area. Large area source, 2.0E+00 s-1 cm-2 to 1.00E+03 s-1 cm-2
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
2 proportional counter
Cs-137: area source between 20 cm2 and 500 cm2, E > 0.5 keV
Reference standard: -particle counting with 2 windowless gas flow counter
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2177
Surface emission rate per unit area. Large area source, 2.0E+00 s-1 cm-2 to 1.00E+03 s-1 cm-2
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
2 proportional counter
Pm-147: area source between 20 cm2 and 500 cm2, E > 0.5 keV
Reference standard: -particle counting with 2 windowless gas flow counter
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2178
Surface emission rate per unit area. Large area source, 2.0E+00 s-1 cm-2 to 1.00E+03 s-1 cm-2
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
2 proportional counter
Tl-204: area source between 20 cm2 and 500 cm2, E > 0.5 keV
Reference standard: -particle counting with 2 windowless gas flow counter
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2179
Surface emission rate per unit area. Large area source, 2.0E+00 s-1 cm-2 to 1.00E+03 s-1 cm-2
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
2 proportional counter
Pb-210/Po-210: area source between 20 cm2 and 500 cm2
Reference standard: -particle counting with 2 windowless gas flow counter
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2180
Surface emission rate per unit area. Large area source, 2.0E+00 s-1 cm-2 to 1.00E+03 s-1 cm-2
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
2 proportional counter
Am-241: area source between 20 cm2 and 500 cm2
Reference standard: -particle counting with 2 windowless gas flow counter
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2181