Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs Ionizing Radiation (2)
IONIZING RADIATION
Air kerma rate (1)
Absorbed dose rate to water (1)
GEOGRAPHIC LOCATION
EURAMET (2)
Croatia (2)
Result of the search
Your query 'Croatia' produced 2 results New search

 
Croatia, IRB-SSDL (Secondary Standard Dosimetry Laboratory of the Ruder Boskovic Institute)
Complete CMCs in Ionizing Radiation for Croatia (.PDF file)
Absorbed dose rate to water. Ionization chamber, 1.6E-03 Gy/s to 2.7E-03 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.8
Calibration against secondary standard in a water phantom
Co-60: IAEA TRS 398 2000
Reference standard: Reference standard ionisation chamber
Source of traceability: IAEA
Approved on 15 July 2017
Internal NMI service identifier: EUR-RAD-IRB-SSDL-1001
Air kerma rate. Ionization chamber, 1.6E-03 Gy/s to 2.7E-03 Gy/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.3
Calibration against secondary standard free in air
Co-60: IAEA TRS 398 2000
Reference standard: Reference standard ionisation chamber
Source of traceability: IAEA
Approved on 15 July 2017
Internal NMI service identifier: EUR-RAD-IRB-SSDL-1002