Result of the search
Your query 'United Kingdom' produced 14 results New search

 
United Kingdom, NMRO (National Measurement and Regulation Office)
Complete CMCs in Mass and related quantities for United Kingdom (.PDF file)
Volume. Micropipettes, 0.02 mL to 5 mL
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1 to 0.09
Gravimetric
Temperature: 20 °C
Delivered volume
Internal NMI service identifier: UK39
 
Volume. Glassware, pipettes, automatic pipettes, etc., 0.002 L to 20 L
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.02 to 0.008
Gravimetric
Temperature: 20 °C
Delivered volume
Internal NMI service identifier: UK40
 
Volume. Metal delivery measures, 1 L to 500 L
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.008
Gravimetric
Temperature: 20 °C
Delivered volume
Internal NMI service identifier: UK41
 
Volume. Metal or glass measures or artefacts, 0.001 L to 500 L
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.02 to 0.008
Gravimetric
Temperature: 20 °C
Contained volume
Internal NMI service identifier: UK42
 
Volume. Metal or glass measures or artefacts, 0.005 L to 20 L
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.02
Volumetric
Temperature: 20 °C
Contained volume
Internal NMI service identifier: UK43
 
Volume. Metal proving tanks, 10 L to 10000 L
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.01
Volumetric
Temperature: 20 °C
Contained volume
Internal NMI service identifier: UK44
 
United Kingdom, TUVNEL (TUVNEL)
Complete CMCs in Mass and related quantities for United Kingdom (.PDF file)
Liquid flowing quantity: mass or volume. Any flow measurement instrument or flow device, 0.002 L/min to 5 L/min
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.05
Pulsed, electrical, digital and optical outputs; various methods
Fluid: kerosine, gas-oil, velocite, water, any safe liquid
Viscosity: up to 30 cSt
Temperature: ambient
Pressure: 0 MPa to 0.5 MPa
Approved on 15 October 2013
Internal NMI service identifier: UK1
 
Liquid flowing quantity: mass or volume. Any flow measurement instrument or flow device, 0.04 L/s to 100 L/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.03
Pulsed, electrical, digital and optical outputs; various methods
Fluid: kerosine, gas-oil, mineral oils
Viscosity: 2.2 cSt to 30 cSt
Temperature: 5 °C to 50 °C
Pressure: 0 MPa to 0.8 MPa
Pipe up to 200 mm
Approved on 15 October 2013
Internal NMI service identifier: UK2
 
Liquid flowing quantity: mass or volume. Any flow measurement instrument or flow device, 0.04 L/s to 100 L/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.05
Pulsed, electrical, digital and optical outputs; various methods
Fluid: heavy oil
Viscosity: 30 cSt and 500 cSt
Temperature: 5 °C to 50 °C
Pressure: 0 MPa to 0.8 MPa
Pipe up to 200 mm
Approved on 15 October 2013
Internal NMI service identifier: UK3
 
Liquid flowing quantity: mass or volume. Any flow measurement instrument or flow device, 0.04 L/s to 200 L/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.08
Pulsed, electrical, digital and optical outputs; various methods
Fluid: kerosine, gas-oil, mineral oils
Viscosity: 2.2 cSt to 30 cSt
Temperature: 5 °C to 50 °C
Pressure: 0 MPa to 0.8 MPa
Pipe up to 200 mm
Approved on 15 October 2013
Internal NMI service identifier: UK4
 
Liquid flowing quantity: mass or volume. Any flow measurement instrument or flow device, 0.04 L/s to 150 L/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.25
Pulsed, electrical, digital and optical outputs; various methods
Fluid: heavy oil
Viscosity: 30 cSt to 500 cSt
Temperature: 5 °C to 50 °C
Pressure: 0 MPa to 0.8 MPa
Pipe up to 200 mm
Approved on 15 October 2013
Internal NMI service identifier: UK5
 
Liquid flowing quantity: mass and volume. Any flow measurement instrument or flow device, 0.1 L/s to 200 L/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.1
Pulsed, electrical, digital and optical outputs; various methods
Fluid: water
Temperature: 15 °C to 40 °C
Pressure: 0 MPa to 1 MPa
Pipe up to 1000 mm
Approved on 15 October 2013
Internal NMI service identifier: UK6
 
Liquid flowing quantity: mass and volume. Any flow measurement instrument or flow device, 0.1 L/s to 400 L/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.15
Pulsed, electrical, digital and optical outputs; various methods
Fluid: water
Temperature: 15 °C to 40 °C
Pressure: 0 MPa to 1 MPa (0.1 l/s to 200 l/s)
Pressure: 0 MPa to 0.5 MPa (200 l/s to 400 l/s)
Pipe up to 1000 mm
Approved on 15 October 2013
Internal NMI service identifier: UK7
 
Volume liquidhydrocarbon flowrate. Any flow measurement instrument or flow device, 0.5 L/s to 100 L/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.08
Pulsed, electrical, digital and optical outputs; various methods
Fluid: Crown oil
Viscosity: 1.5 cP to 10 cP
Temperature: 20 °C to 80 °C
Pressure: 0.5 MPa to 9.4 MPa
Also Mass hydrocarbon flow rate
Approved on 17 November 2017
Internal NMI service identifier: UK8