Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (56)
PHYSICS
Materials
Temperature (56)
Temperature (56)
GEOGRAPHIC LOCATION
EURAMET (56)
United Kingdom (56)
Result of the search
Your query 'United Kingdom' produced 56 results New search
1  2  3

 
United Kingdom, NPL (National Physical Laboratory)
Complete CMCs in Thermometry for United Kingdom (.PDF file)
Temperature. Cell for contact thermometry, 83.8058 K
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.00025
Triple point of Argon for capsule SPRT
Approved on 18 May 2004
Internal NMI service identifier: NPL/PK01
 
Temperature. Cell for contact thermometry, 156.5985 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.0007
Freezing point of Indium
Approved on 18 May 2004
Internal NMI service identifier: NPL/PK01
 
Temperature. Cell for contact thermometry, 231.928 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.0006
Freezing point of Tin
Approved on 18 May 2004
Internal NMI service identifier: NPL/PK01
 
Temperature. Cell for contact thermometry, 419.527 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.0009
Freezing point of Zinc
Approved on 18 May 2004
Internal NMI service identifier: NPL/PK01
 
Temperature. Long-stem SPRT, 156.5985 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.0007
Calibration at Indium freezing point
Approved on 18 May 2004
Internal NMI service identifier: NPL/PM02
 
Temperature. Long-stem SPRT, 231.928 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.0007
Calibration at Tin freezing point
Approved on 18 May 2004
Internal NMI service identifier: NPL/PM02
 
Temperature. Long-stem SPRT, 419.527 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.0009
Calibration at Zinc freezing point
Approved on 18 May 2004
Internal NMI service identifier: NPL/PM02
 
Temperature. Long-stem SPRT, 660.323 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.0025
Calibration at Aluminium freezing point
Approved on 18 May 2004
Internal NMI service identifier: NPL/PM02
 
Temperature. Long-stem SPRT, 961.78 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.004
Calibration at Silver freezing point
Approved on 18 May 2004
Internal NMI service identifier: NPL/PM02
 
Temperature. Platinum Resistance Thermometer, -100 °C to -80 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.01
Calibration by comparison
Bath: acetone
Approved on 03 November 2009
Internal NMI service identifier: NPL/PM04
 
Temperature. Noble-metal thermocouple (type R, S), 419.527 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.13
Calibration at fixed point
Approved on 03 November 2009
Internal NMI service identifier: NPL/PM03
 
Temperature. Noble-metal thermocouple (type R, S), 961.78 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.21
Calibration at fixed point
Approved on 03 November 2009
Internal NMI service identifier: NPL/PM03
 
Temperature. Noble-metal thermocouple (type R, S), 1084.62 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.21
Calibration at fixed point
Approved on 03 November 2009
Internal NMI service identifier: NPL/PM03
 
Temperature. Noble-metal thermocouple (type R, S), 1324.3 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.53
Calibration at fixed point
Approved on 03 November 2009
Internal NMI service identifier: NPL/PM03
 
Temperature. Noble-metal thermocouple (type B)), 419.527 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.27
Calibration at fixed point
Approved on 03 November 2009
Internal NMI service identifier: NPL/PM03
 
Temperature. Noble-metal thermocouple (type B)), 961.78 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.22
Calibration at fixed point
Approved on 03 November 2009
Internal NMI service identifier: NPL/PM03
 
Temperature. Noble-metal thermocouple (type B)), 1084.62 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.22
Calibration at fixed point
Approved on 03 November 2009
Internal NMI service identifier: NPL/PM03
 
Temperature. Noble-metal thermocouple (Pt/Pd), 419.527 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.1
Calibration at fixed point
Approved on 03 November 2009
Internal NMI service identifier: NPL/PM03
 
Temperature. Noble-metal thermocouple (Pt/Pd), 961.78 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.07
Calibration at fixed point
Approved on 03 November 2009
Internal NMI service identifier: NPL/PM03
 
Temperature. Noble-metal thermocouple (Pt/Pd), 1084.62 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.07
Calibration at fixed point
Approved on 03 November 2009
Internal NMI service identifier: NPL/PM03