Croatia, FSB-LPM (FSB - Laboratory for Process Measurements)
Complete CMCs in Thermometry for Croatia (.PDF file)
Temperature. Platinum Resistance Thermometer, 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 8
Comparison with SPRT
Ice point
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.2.2
 
Temperature. Platinum Resistance Thermometer, -40 °C to 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 10
Comparison with SPRT
Glycol bath
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.2.2
 
Temperature. Platinum Resistance Thermometer, 0 °C to 95 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 10
Comparison with SPRT
Water bath
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.2.2
 
Temperature. Platinum Resistance Thermometer, 90 °C to 180 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 12
Comparison with SPRT
Oil bath
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.2.2
 
Temperature. Platinum Resistance Thermometer, 180 °C to 660 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 100
Comparison with SPRT
Fluidised bath
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.2.2
 
Temperature. Noble metal thermocouple, 1084.62 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.6
Calibration at Cu freezing point
3-zone furnace
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.1 Cu
 
Temperature. Noble metal thermocouple, 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.20
Secondary fixed point
Ice point
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.1
 
Temperature. Noble metal thermocouple, -40 °C to 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.25
Comparison with SPRT
Glycol bath
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.1
 
Temperature. Noble metal thermocouple, 0 °C to 95 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.25
Comparison with SPRT
Water bath
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.1
 
Temperature. Noble metal thermocouple, 90 °C to 180 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.3
Comparison with SPRT
Oil bath
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.1
 
Temperature. Noble metal thermocouple, 180 °C to 660 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.6
Comparison with SPRT
Fluidised bath
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.1
 
Temperature. Noble metal thermocouple, 660 °C to 1050 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 1.2
Comparison with standard thermocouple
Tube furnace
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.1
 
Temperature. Base metal thermocouple, 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.2
Secondary fixed point
Ice point
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.2
 
Temperature. Base metal thermocouple, -40 °C to 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.2
Comparison with SPRT
Glycol bath
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.2
 
Temperature. Base metal thermocouple, 0 °C to 95 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.2
Comparison with SPRT
Water bath
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.2
 
Temperature. Base metal thermocouple, 90 °C to 180 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.3
Comparison with SPRT
Oil bath
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.2
 
Temperature. Base metal thermocouple, 180 °C to 660 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.8
Comparison with SPRT
Fluidised bath
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.2
 
Temperature. Base metal thermocouple, 660 °C to 1050 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 2
Comparison with standard thermocouple
Tube furnace
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.2
 
Temperature. Pure metal thermocouple, Pt Pd, 1084.62 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.2
Calibration at Cu freezing point
3-zone furnace
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.3.3 Cu
 
Temperature. Temperature sensor with display unit, 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 8
Secondary fixed point
Ice point
Approved on 03 November 2009
Internal NMI service identifier: FSB-LPM/2.7