Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs Ionizing Radiation (2)
IONIZING RADIATION
Reference air kerma rate (2)
GEOGRAPHIC LOCATION
APMP (2)
Korea, Republic of (2)
Result of the search
Your query 'kriss' produced 2 results New search

 
Korea, Republic of, KRISS (Korea Research Institute of Standards and Science)
Complete CMCs in Ionizing Radiation for Korea, Republic of (.PDF file)
Reference air kerma rate. Reference radiation field, 8.8E-05 Gy/h to 8.8E+00 Gy/h
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1
Calibration with a transfer standard
Cs-137: 1 m to 4.5 m from the source
Reference standard: Calibrated ionization chamber
Source of traceability: KRISS
Approved on 01 June 2011
Internal NMI service identifier: APM-RAD-KRISS-1005
Reference air kerma rate. Reference radiation field, 4.4E-06 Gy/h to 8.8E-05 Gy/h
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
Calibration with a transfer standard
Cs-137: 1 m to 4.5 m from the source
Reference standard: Calibrated ionization chamber
Source of traceability: KRISS
Approved on 01 June 2011
Internal NMI service identifier: APM-RAD-KRISS-1006