Result of the search
Your query 'United Kingdom' produced 11 results New search

 
United Kingdom, NPL (National Physical Laboratory)
Complete CMCs in Acoustics, Ultrasound, Vibration for United Kingdom (.PDF file)
Free-field sensitivity level. Reference measuring hydrophones/ projectors. Measurand unit, V/µPa
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.7
Three-transducer spherical wave reciprocity in a laboratory tank, IEC 565:1977
Frequency: 1 kHz to 2 kHz
Approved on 29 September 2004
Internal NMI service identifier: NPL/AW/10/01
 
Free-field sensitivity level. Reference measuring hydrophones/ projectors. Measurand unit, V/µPa
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.5
Three-transducer spherical wave reciprocity in a laboratory tank; IEC 565:1977
Frequency: 2 kHz to 500 kHz
Approved on 29 September 2004
Internal NMI service identifier: NPL/AW/10/01
 
Free-field sensitivity level. Reference measuring hydrophones. Measurand unit, V/µPa
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.9
Comparison with reference hydrophone, IEC 565:1977
Frequency: 1 kHz to 2 kHz
Approved on 29 September 2004
Internal NMI service identifier: NPL/AW/10/02
 
Free-field sensitivity level. Reference measuring hydrophones. Measurand unit, V/µPa
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.7
Comparison with reference hydrophone, IEC 565:1977
Frequency: 2 kHz to 1 MHz
Approved on 29 September 2004
Internal NMI service identifier: NPL/AW/10/02
 
Pressure sensitivity level. Reference measuring hydrophones. Measurand unit, V/µPa
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.5
Comparison with a calibrated reference microphone
Frequency: 25 Hz to 400 Hz
Approved on 29 September 2004
Internal NMI service identifier: NPL/AW/10/03
 
Free-field sensitivity. Hydrophone. Measurand unit, µV/Pa
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 6
Comparison with a secondary hydrophone in a non-linearly distorted sound field
Frequency: 1 MHz to 8 MHz
Approved on 29 September 2004
Internal NMI service identifier: NPL/AW/07/01
 
Free-field sensitivity. Hydrophone. Measurand unit, µV/Pa
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 7
Comparison with a secondary hydrophone in a non-linearly distorted sound field
Frequency: 9 MHz to 12 MHz
Approved on 29 September 2004
Internal NMI service identifier: NPL/AW/07/01
 
Free-field sensitivity. Hydrophone. Measurand unit, µV/Pa
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 8
Comparison with a secondary hydrophone in a non-linearly distorted sound field
Frequency: 13 MHz to 16 MHz
Approved on 29 September 2004
Internal NMI service identifier: NPL/AW/07/01
 
Free-field sensitivity. Hydrophone. Measurand unit, µV/Pa
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 11
Comparison with a secondary hydrophone in a non-linearly distorted sound field
Frequency: 17 MHz to 20 MHz
Approved on 29 September 2004
Internal NMI service identifier: NPL/AW/07/01
 
Ultrasonic power. Plane piston ultrasound transducer and generator, 0.0005 W to 1 W
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.5 to 6
Primary standard radiation force balance
Frequency: 0.8 MHz to 20 MHz
Approved on 29 September 2004
Internal NMI service identifier: NPL/AW/08/05
 
Ultrasonic power. Plane piston ultrasound transducer and generator, 0.50 W to 20 W
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.5 to 5
Calibrated radiation force balance
Frequency: 0.8 MHz to 3 MHz
Approved on 29 September 2004
Internal NMI service identifier: NPL/AW/08/05