Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (65)
PHYSICS >
Photometry
Properties of detectors and sources
Properties of materials (65)
Absorbance, regular, spectral (20)
Transmittance, regular, spectral (12)
Colour, surface, x, y, Y (6)
Colour, surface, L*a*b* (6)
Reflectance, regular, spectral (3)
Transmittance, diffuse, spectral (2)
Colour, transmitted, x, y, Y (2)
Emittance, spectral (2)
BRDF (2)
Reflectance factor (2)
Radiance factor (2)
Colour, transmitted, L*a*b* (2)
IONIZING RADIATION
Emission rate
GEOGRAPHIC LOCATION
EURAMET (65)
United Kingdom (65)
Result of the search
Your query 'United Kingdom' produced 65 results New search
[Prev] 2  3  4

 
United Kingdom, NPL (National Physical Laboratory)
Complete CMCs in Photometry and Radiometry for United Kingdom (.PDF file)
Colour, surface, L*a*b*. Fluorescent material, L* = 0 to L* = 160
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.003
Reference spectrofluorimeter
Irradiation conditions: monochromator
Specific measurement conditions: 0/45
Type of fluorescent material: general
Approved on 07 May 2007
 
Colour, surface, L*a*b*. Fluorescent material, a*, b* = -200 to a*, b* = 200
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 1.2
Reference spectrofluorimeter
Irradiation conditions: monochromator
Specific measurement conditions: 0/45
Type of fluorescent material: general
Approved on 07 May 2007
 
Colour, surface, L*a*b*. Diffusely reflecting material, L* = 0 to L* = 100
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.15
Integrating sphere spectrophotometer
Specific measurement conditions: 0/d, 2 nm bandwidth
Other types of material can also be calibrated
Approved on 07 May 2007
 
Colour, surface, L*a*b*. Diffusely reflecting material, a*, b* = -200 to a*, b* = 200
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.1
Integrating sphere spectrophotometer
Specific measurement conditions: 0/d, 2 nm bandwidth
Other types of material can also be calibrated
Approved on 07 May 2007
 
Colour, transmitted, x, y, Y. General material, x, y = 0 to x, y = 0.9
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.0002
Spectrophotometer
Specific measurement conditions: regular transmittance geometry
Type of material: non-fluorescent
 
Colour, transmitted, x, y, Y. General material, Y = 0 to Y = 1
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.003
Spectrophotometer
Specific measurement conditions: regular transmittance geometry
Type of material: non-fluorescent
 
Colour, transmitted, L*a*b*. General material, L* = 0 to L* = 100
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.05
Spectrophotometer
Specific measurement conditions: regular transmittance geometry
Type of material: non-fluorescent
 
Colour, transmitted, L*a*b*. General material, a*, b* = -200 to a*, b* = 200
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.05
Spectrophotometer
Specific measurement conditions: regular transmittance geometry
Type of material: non-fluorescent
 
Transmittance, regular, spectral. Spectrally-neutral material, 1E-03 to 1
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 3E-05 to 3.5E-04, varies as the logarithm of the measurand
Reference spectrophotometer
Wavelength: 250 nm to 749 nm
Bandwidth: < 3 nm
 
Transmittance, regular, spectral. Spectrally-neutral material, 1E-03 to 1
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 1E-05 to 4.5E-04, varies as the logarithm of the measurand
Reference spectrophotometer
Wavelength: 750 nm to 2500 nm
Bandwidth: < 3 nm
 
Transmittance, regular, spectral. Spectrally-neutral material, 1E-03 to 1
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.0017 at 0.73
Scanning spectrophotometer
Wavenumber: 3990 cm-1
Bandwidth: 2 cm-1 to 15 cm-1
Specific measurement conditions: measured at 5 ° and f/6. Values calculated for normal incidence and a range of f/n's.
Other types of material can also be calibrated. Uncertainties at other transmittance values available on request.
Approved on 07 May 2007
 
Transmittance, regular, spectral. Spectrally-neutral material, 1E-03 to 1
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.00065 at 0.14
Scanning spectrophotometer
Wavenumber: 3512 cm-1
Bandwidth: 2 cm-1 to 15 cm-1
Specific measurement conditions: measured at 5 ° and f/6. Values calculated for normal incidence and a range of f/n's.
Other types of material can also be calibrated. Uncertainties at other transmittance values available on request.
Approved on 07 May 2007
 
Transmittance, regular, spectral. Spectrally-neutral material, 1E-03 to 1
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.0011 at 0.36
Scanning spectrophotometer
Wavenumber: 3031 cm-1
Bandwidth: 2 cm-1 to 15 cm-1
Specific measurement conditions: measured at 5 ° and f/6. Values calculated for normal incidence and a range of f/n's.
Other types of material can also be calibrated. Uncertainties at other transmittance values available on request.
Approved on 07 May 2007
 
Transmittance, regular, spectral. Spectrally-neutral material, 1E-03 to 1
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.00052 at 0.06
Scanning spectrophotometer
Wavenumber: 2739 cm-1
Bandwidth: 2 cm-1 to 15 cm-1
Specific measurement conditions: measured at 5 ° and f/6. Values calculated for normal incidence and a range of f/n's.
Other types of material can also be calibrated. Uncertainties at other transmittance values available on request.
Approved on 07 May 2007
 
Transmittance, regular, spectral. Spectrally-neutral material, 1E-03 to 1
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.00079 at 0.17
Scanning spectrophotometer
Wavenumber: 2598 cm-1
Bandwidth: 2 cm-1 to 15 cm-1
Specific measurement conditions: measured at 5 ° and f/6. Values calculated for normal incidence and a range of f/n's.
Other types of material can also be calibrated. Uncertainties at other transmittance values available on request.
Approved on 07 May 2007
 
Transmittance, regular, spectral. Spectrally-neutral material, 1E-03 to 1
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.00064 at 0.10
Scanning spectrophotometer
Wavenumber: 2473 cm-1
Bandwidth: 2 cm-1 to 15 cm-1
Specific measurement conditions: measured at 5 ° and f/6. Values calculated for normal incidence and a range of f/n's.
Other types of material can also be calibrated. Uncertainties at other transmittance values available on request.
Approved on 07 May 2007
 
Transmittance, regular, spectral. Spectrally-neutral material, 1E-03 to 1
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.0001 at 0.0001
Scanning spectrophotometer
Wavenumber: 2010 cm-1
Bandwidth: 2 cm-1 to 15 cm-1
Specific measurement conditions: measured at 5 ° and f/6. Values calculated for normal incidence and a range of f/n's.
Other types of material can also be calibrated. Uncertainties at other transmittance values available on request.
Approved on 07 May 2007
 
Transmittance, diffuse, spectral. Spectrally-neutral material, 0 to 1
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.01
Spectrophotometer with integrating sphere accessory
Wavelength range: 200 nm to 800 nm
Bandwidth: 0.5 nm to 2 nm
Specific measurement conditions: 0/d
Other types of material can also be calibrated
 
Transmittance, diffuse, spectral. Spectrally-neutral material, 0 to 1
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.01
Spectrophotometer with integrating sphere accessory
Wavelength range: 800 nm to 2500 nm
Bandwidth: 2 nm to 20 nm
Specific measurement conditions: 0/d
Other types of material can also be calibrated
 
Absorbance, regular, spectral. Spectrally-neutral material, 0.05 to 0.22
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.0008 to 0.0007
Reference spectrophotometer
Wavelength range: 200 nm to 209 nm
Bandwidth: < 4 nm
Specific measurement conditions: CIE geometries
Other types of material can also be calibrated
Approved on 07 May 2007