| Metrologia, 2012, 49(3), 257-265 |
| Investigation of low-temperature fixed points by an international star intercomparison of sealed triple-point cells |
| Fellmuth B., Wolber L., Head D.I., Hermier Y., Hill K.D., Nakano T., Pavese F., Peruzzi A., Rusby R.L., Shkraba V., Steele A.G., Steur P.P.M., Szmyrka-Grzebyk A., Tew W.L., Wang L., White D.R. |
 |
Pavese F. Critical review of information relevant to the correction of the effect of chemical impurities in gases used for the realization of ITS-90 fixed points
Metrologia, 2009, 46, n°1, 47-61
|
 |
Fellmuth B., Wolber L., Hermier Y., Pavese F., Steur P.P.M., Peroni I., Szmyrka-Grzebyk A., Lipinski L., Tew W.L., Nakano T., Sakurai H., Tamura O., Head D., Hill K.D., Steele A.G. Isotopic and other influences on the realization of the triple point of hydrogen
Metrologia, 2005, 42, n°4, 171-193
|
 |
Steele A.G., Fellmuth B., Head D.I., Hermier Y., Kang K.H., Steur P.P.M., Tew W.L. Key Comparison: CCT-K2: Key comparison of capsule-type standard platinum resistance thermometers from 13.8 K to 273.16 K
Metrologia, 2002, 39, n°6, 551-571
|
 |
Ancsin J. Development of Cryogenic Sealed Cells as Temperature Standards: their Filling System and the Purification of CO2
Metrologia, 1992, 29, n°1, 71-78
|
 |
Pavese F., Ancsin J., Astrov D.N., Bonhoure J., Bonnier G., Furukawa G.T., Kemp R.C., Maas H., Rusby R.L., Sakurai H., Ling Shan-Kang An International Intercomparison of Fixed Points by Means of Sealed Cells in the Range 13.81 K - 90.686 K
Metrologia, 1984, 20, n°4, 127-144
|
|
|