Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (16)
PHYSICS
Fluid flow
Force
Torque, Viscosity, Hardness and Gravity
Impedance up to the MHz range (16)
Capacitance: low loss capacitors (4)
Capacitance: dielectric capacitors (4)
Inductance: self inductance, intermediate values (4)
Capacitance: meters (1)
Inductance: self inductance, high values (1)
Inductance: meters (1)
Inductance: self inductance, low values (1)
CHEMICAL ANALYTE
lead
GEOGRAPHIC LOCATION
SIM (16)
Mexico (16)
Result of the search
Your query 'cenam' produced 16 results New search

 
Mexico, CENAM (Centro Nacional de Metrologia)
Complete CMCs in Electricity and Magnetism for Mexico (.PDF file)
Capacitance: meters. Capacitance bridge, LCR meter (capacitance), 0.001 nF to 1000 nF
Relative expanded uncertainty (k = 2, level of confidence 95%) in µF/F: 0.5 to 40
Direct measurement
Frequency: 1 kHz
Voltage: 1 mV to 50 V
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-004B
Uncertainty table
 
Inductance: meters. LCR meter (inductance), 0.1 mH to 10000 mH
Relative expanded uncertainty (k = 2, level of confidence 95%) in mH/H: 0.05 to 0.63
Direct measurement
Frequency: 1 kHz
Current: 10 µA to 30 mA
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-004B
Uncertainty table
 
Inductance: self inductance, intermediate values. Fixed inductor, 100 mH
Relative expanded uncertainty (k = 2, level of confidence 95%) in µH/H: 90
Substitution using an LCR meter
Frequency: 1 kHz
Current: 1 mA
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-010-04
 
Inductance: self inductance, intermediate values. Fixed inductor, 1 H
Relative expanded uncertainty (k = 2, level of confidence 95%) in mH/H: 0.11
Substitution using an LCR meter
Frequency: 1 kHz
Current: 100 µA
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-010-05
 
Inductance: self inductance, high values. Fixed inductor, 10 H
Relative expanded uncertainty (k = 2, level of confidence 95%) in mH/H: 0.16
Substitution using an LCR meter
Frequency: 1 kHz
Current: 10 µA
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-010-06
 
Inductance: self inductance, intermediate values. Fixed inductor, 1 mH
Relative expanded uncertainty (k = 2, level of confidence 95%) in mH/H: 0.10
Substitution using an LCR meter
Frequency: 1 kHz
Current: 30 mA
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-010-02
 
Inductance: self inductance, intermediate values. Fixed inductor, 10 mH
Relative expanded uncertainty (k = 2, level of confidence 95%) in µH/H: 50
Substitution using an LCR meter
Frequency: 1 kHz
Current: 6 mA
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-010-03
 
Capacitance: low loss capacitors. Standard capacitor (fused silica dielectric), 1 pF
Relative expanded uncertainty (k = 2, level of confidence 95%) in µF/F: 1
Comparison using a capacitance bridge based on inductive voltage dividers
Frequency: 1 kHz
Voltage: 50 V
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-001-01
 
Capacitance: low loss capacitors. Standard capacitor (fused silica dielectric), 10 pF
Relative expanded uncertainty (k = 2, level of confidence 95%) in µF/F: 0.5
Comparison using a capacitance bridge based on inductive voltage dividers
Frequency: 1 kHz
Voltage: 50 V
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-001-02
 
Capacitance: low loss capacitors. Standard capacitor (fused silica dielectric), 100 pF
Relative expanded uncertainty (k = 2, level of confidence 95%) in µF/F: 0.6
Comparison using a capacitance bridge based on inductive voltage dividers
Frequency: 1 kHz
Voltage: 50 V
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-001-03
 
Capacitance: low loss capacitors. Standard capacitor (dry-nitrogen), 1 nF
Relative expanded uncertainty (k = 2, level of confidence 95%) in µF/F: 1.5
Substitution using an 8-digits capacitance meter
Frequency: 1 kHz
Voltage: 0.75 V
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-002-01
 
Capacitance: dielectric capacitors. Fixed capacitor (mica), 1 nF
Relative expanded uncertainty (k = 2, level of confidence 95%) in µF/F: 10
Substitution using an 8-digits capacitance meter
Frequency: 1 kHz
Voltage: 0.75 V
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-004-01
 
Capacitance: dielectric capacitors. Fixed capacitor (mica), 10 nF
Relative expanded uncertainty (k = 2, level of confidence 95%) in µF/F: 35
Substitution using an 8-digits capacitance meter
Frequency: 1 kHz
Voltage: 100 mV
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-004-02
 
Capacitance: dielectric capacitors. Fixed capacitor (mica), 100 nF
Relative expanded uncertainty (k = 2, level of confidence 95%) in µF/F: 40
Substitution using an 8-digits capacitance meter
Frequency: 1 kHz
Voltage: 10 mV
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-004-03
 
Capacitance: dielectric capacitors. Fixed capacitor (mica), 1 µF
Relative expanded uncertainty (k = 2, level of confidence 95%) in µF/F: 45
Substitution using an 8-digits capacitance meter
Frequency: 1 kHz
Voltage: 1 mV
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-004-04
 
Inductance: self inductance, low values. Fixed inductor, 100 µH
Relative expanded uncertainty (k = 2, level of confidence 95%) in mH/H: 0.60
Substitution using an LCR meter
Frequency: 1 kHz
Current: 30 mA
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-010-01