Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (2)
PHYSICS
Pressure (2)
Gauge pressure, gas medium (2)
Gauge pressure, gas medium
GEOGRAPHIC LOCATION
APMP (2)
Thailand (2)
Result of the search
Your query 'NIMT' produced 2 results New search

 
Thailand, NIMT (National Institute of Metrology (Thailand))
Complete CMCs in Mass and related quantities for Thailand (.PDF file)
Gauge pressure, gas medium. Pressure balance, 3.5 kPa to 1750 kPa
Absolute expanded uncertainty (k = 2, level of confidence 95%) in kPa: 3E-05P, P pressure in kPa, or 0.8 Pa whichever is greater
Direct comparison with pressure standard: crossfloat
Approved on 18 April 2006.
Uncertainty values range from 0.8 Pa to 52.5 Pa
Internal NMI service identifier: NIMT/26
 
Gauge pressure, gas medium. Pressure balance, 1.75 MPa to 7 MPa
Absolute expanded uncertainty (k = 2, level of confidence 95%) in MPa: 4E-05P, P pressure in MPa
Direct comparison with pressure standard: crossfloat
Approved on 18 April 2006.
Uncertainty values range from 70 Pa to 280 Pa
Internal NMI service identifier: NIMT/27