Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (29)
PHYSICS
Mass, mass standards
DC voltage, current, and resistance (29)
DC resistance sources: intermediate values (higher than 1 ohm to 1 Mohm) (6)
DC resistance sources: high values (higher than 1 Mohm) (6)
DC voltage sources: intermediate values (higher than 10 V to 1100 V) (4)
DC resistance sources: low values (equal or below 1 ohm) (4)
DC current sources: intermediate values (higher than 0.1 mA to 20 A) (3)
DC voltage sources: low values (equal or below 10 V) (2)
DC current sources: low values (equal or below 0.1 mA) (2)
DC voltage sources: low value ranges (equal or below 10 V) (1)
DC current sources: high values (higher than 20 A to 100 A) (1)
DC voltage sources: single values
CHEMICAL ANALYTE
carbon monoxide
GEOGRAPHIC LOCATION
APMP (29)
Thailand (29)
Result of the search
Your query 'NIMT' produced 29 results New search
1  2

 
Thailand, NIMT (National Institute of Metrology (Thailand))
Complete CMCs in Electricity and Magnetism for Thailand (.PDF file)
DC voltage sources: low value ranges (equal or below 10 V). DC voltage source, multifunction calibrator, voltage V, 0 V to 10 V
Absolute expanded uncertainty (k = 2, level of confidence 95%) in µV: (0.5 v + 2), v in V
Compare against DC standard
Internal NMI service identifier: NIMT/7
 
DC voltage sources: intermediate values (higher than 10 V to 1100 V). DC voltage source, multifunction calibrator, voltage V, > 10 V to 100 V
Absolute expanded uncertainty (k = 2, level of confidence 95%) in µV: (0.5 v + 15), v in V
Compare against DC standard
Internal NMI service identifier: NIMT/8
 
DC voltage sources: intermediate values (higher than 10 V to 1100 V). DC voltage source, multifunction calibrator, voltage V, > 100 V to 1000 V
Absolute expanded uncertainty (k = 2, level of confidence 95%) in µV: (0.8 v + 200), v in V
Compare against DC standard
Internal NMI service identifier: NIMT/9
 
DC voltage sources: low values (equal or below 10 V). DC voltage source, multifunction calibrator, 100 mV
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 5
Compare against DC standard
Internal NMI service identifier: NIMT/3
 
DC voltage sources: low values (equal or below 10 V). DC voltage source, multifunction calibrator, 1 V
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.6
Compare against DC standard
Internal NMI service identifier: NIMT/4
 
DC voltage sources: intermediate values (higher than 10 V to 1100 V). DC voltage source, multifunction calibrator, 100 V
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.5
Compare against DC standard
Internal NMI service identifier: NIMT/5
 
DC voltage sources: intermediate values (higher than 10 V to 1100 V). DC voltage source, multifunction calibrator, 1000 V
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.7
Compare against DC standard
Internal NMI service identifier: NIMT/6
 
DC resistance sources: low values (equal or below 1 ohm). Fixed resistor, 0.001
Relative expanded uncertainty (k = 2, level of confidence 95%) in µ/: 2.8
DCC bridge & range extender
Current: 10 A to 100 A
Oil bath temperature: 23 °C
Internal NMI service identifier: NIMT/10
 
DC resistance sources: low values (equal or below 1 ohm). Fixed resistor, 0.01
Relative expanded uncertainty (k = 2, level of confidence 95%) in µ/: 2.3
DCC bridge & range extender
Current: 1 A to 10 A
Oil bath temperature: 23 °C
Internal NMI service identifier: NIMT/11
 
DC resistance sources: low values (equal or below 1 ohm). Fixed resistor, 0.1
Relative expanded uncertainty (k = 2, level of confidence 95%) in µ/: 2.3
DCC bridge & range extender
Current: 0.1 A to 1 A
Oil bath temperature: 23 °C
Internal NMI service identifier: NIMT/12
 
DC resistance sources: low values (equal or below 1 ohm). Fixed resistor, 1
Relative expanded uncertainty (k = 2, level of confidence 95%) in µ/: 1.5
DCC bridge
Current: 100 A
Oil bath temperature: 23 °C
Internal NMI service identifier: NIMT/13
 
DC resistance sources: intermediate values (higher than 1 ohm to 1 Mohm). Fixed resistor, 10
Relative expanded uncertainty (k = 2, level of confidence 95%) in µ/: 1.6
DCC bridge
Current: 30 mA
Oil bath temperature: 23 °C
Internal NMI service identifier: NIMT/14
 
DC resistance sources: intermediate values (higher than 1 ohm to 1 Mohm). Fixed resistor, 100
Relative expanded uncertainty (k = 2, level of confidence 95%) in µ/: 2.2
DCC bridge
Current: 10 mA
Oil bath temperature: 23 °C
Internal NMI service identifier: NIMT/15
 
DC resistance sources: intermediate values (higher than 1 ohm to 1 Mohm). Fixed resistor, 1 k
Relative expanded uncertainty (k = 2, level of confidence 95%) in µ/: 2.5
DCC bridge
Current: 3 mA
Oil bath temperature: 23 °C
Internal NMI service identifier: NIMT/16
 
DC resistance sources: intermediate values (higher than 1 ohm to 1 Mohm). Fixed resistor, 10 k
Relative expanded uncertainty (k = 2, level of confidence 95%) in µ/: 2.4
DCC bridge
Current: 1 mA
Oil bath temperature: 23 °C
Internal NMI service identifier: NIMT/17
 
DC resistance sources: intermediate values (higher than 1 ohm to 1 Mohm). Fixed resistor, 100 k
Relative expanded uncertainty (k = 2, level of confidence 95%) in µ/: 2.5
DCC bridge
Current: 0.1 mA
Oil bath temperature: 23 °C
Internal NMI service identifier: NIMT/18
 
DC resistance sources: intermediate values (higher than 1 ohm to 1 Mohm). Fixed resistor, 1 M
Relative expanded uncertainty (k = 2, level of confidence 95%) in µ/: 18
Wheastone bridge
Voltage: 1 V
Internal NMI service identifier: NIMT/19
 
DC resistance sources: high values (higher than 1 Mohm). Fixed resistor, 10 M
Relative expanded uncertainty (k = 2, level of confidence 95%) in µ/: 32
Wheastone bridge
Voltage: 1 V
Internal NMI service identifier: NIMT/20
 
DC resistance sources: high values (higher than 1 Mohm). Fixed resistor, 100 M
Relative expanded uncertainty (k = 2, level of confidence 95%) in m/: 0.12
Wheastone bridge
Voltage: 5 V
Internal NMI service identifier: NIMT/21
 
DC resistance sources: high values (higher than 1 Mohm). Fixed resistor, 1 G
Relative expanded uncertainty (k = 2, level of confidence 95%) in m/: 0.69
Wheastone bridge
Voltage: 20 V
Internal NMI service identifier: NIMT/22