Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (8)
PHYSICS
Dimensional metrology
Properties of detectors and sources (8)
Color, emitted, x, y (1)
Correlated color temperature (1)
Irradiance, spectral (2)
Responsivity, laser, power (1)
Responsivity, spectral, power (3)
GEOGRAPHIC LOCATION
SIM (8)
Mexico (8)
Result of the search
Your query 'cenam' produced 8 results New search

 
Mexico, CENAM (Centro Nacional de Metrologia)
Complete CMCs in Photometry and Radiometry for Mexico (.PDF file)
Correlated color temperature. Tungsten lamp, 2000 K to 3200 K
Absolute expanded uncertainty (k = 2, level of confidence ~95%) in K: 50 to 100, varies with measurand
Spectroradiometer
Internal NMI service identifier: CENAM/520-D005-003
 
Color, emitted, x, y. General source, no range given
Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.006
Spectroradiometer
Bandwidth: 2.5 nm
Type of source: tungsten lamp
Internal NMI service identifier: CENAM/520-D005-003
 
Responsivity, laser, power. General detector. Measurand unit, A/W or V/W or Reading/W
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 0.03
Laser source
Wavelength: 633 nm
Power level: 0.1 mW
Type of detector: Si trap
Si, ECPR detectors can be measured, uncertainty evaluated on each case
Internal NMI service identifier: CENAM/520-D007-001
 
Responsivity, spectral, power. Broad band detector. Measurand unit, A/W, V/W, Reading/W
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 0.8 to 0.62, varies with wavelength
Monochromator plus tungsten source
Wavelength range: 350 nm to 450 nm
Bandwidth: 4 nm
Power level: 0.4 µW to 4 µW
Internal NMI service identifier: CENAM/520-D007-001
 
Responsivity, spectral, power. Broad band detector. Measurand unit, A/W, V/W, Reading/W
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 0.62 to 0.76, varies with wavelength
Monochromator plus tungsten source
Wavelength range: 450 nm to 770 nm
Bandwidth: 4 nm
Power level: 0.4 µW to 4 µW
Internal NMI service identifier: CENAM/520-D007-001
 
Responsivity, spectral, power. Broad band detector. Measurand unit, A/W, V/W, Reading/W
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 0.76 to 1.5, varies with wavelength
Monochromator plus tungsten source
Wavelength range: 770 nm to 1100 nm
Bandwidth: 4 nm
Power level: 0.4 µW to 4 µW
Internal NMI service identifier: CENAM/520-D007-001
 
Irradiance, spectral. Tungsten lamp, 7.70E-03 (W/m2)/nm to 1.09E-01 (W/m2)/nm
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 2.5
Spectroradiometer
Wavelength range: 400 nm to 795 nm
Bandwidth: 2.5 nm
Internal NMI service identifier: CENAM/520-D005-007
 
Irradiance, spectral. Tungsten lamp, 1.11E-01 (W/m2)/nm to 1.13E-01 (W/m2)/nm
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 4
Spectroradiometer
Wavelength range: 800 nm to 995 nm
Bandwidth: 2.5 nm
Internal NMI service identifier: CENAM/520-D005-007