Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs Chemistry
CMCs General Physics (373)
PHYSICS
Impedance up to the MHz range
Temperature
Frequency
AC voltage, current, and power (254)
Dimensional metrology (32)
DC voltage, current, and resistance (31)
Mass, mass standards (19)
Sound in air (18)
Pressure (11)
Properties of detectors and sources (5)
Laser frequencies (2)
Time interval (1)
GEOGRAPHIC LOCATION
APMP (373)
Thailand (373)
Result of the search
Your query 'NIMT' produced 373 results New search
1  2  3 [Next >>]

 
Thailand, NIMT (National Institute of Metrology (Thailand))
Complete CMCs in Acoustics, Ultrasound, Vibration for Thailand (.PDF file)
Pressure sensitivity level. Measurement microphone type LS1P. Measurand unit, dB (reference: 1 V/Pa)
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB (reference: 1 V/Pa): 0.05
IEC 61094-2:1992
Frequency: 20 Hz
Approved on 11 January 2013
Internal NMI service identifier: NIMT/14010 - 11101
 
Pressure sensitivity level. Measurement microphone type LS1P. Measurand unit, dB (reference: 1 V/Pa)
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB (reference: 1 V/Pa): 0.04
IEC 61094-2:1992
Frequency: 31.5 Hz to 4 kHz
Approved on 11 January 2013
Internal NMI service identifier: NIMT/14010 - 11101
 
Pressure sensitivity level. Measurement microphone type LS1P. Measurand unit, dB (reference: 1 V/Pa)
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB (reference: 1 V/Pa): 0.06
IEC 61094-2:1992
Frequency: 8 kHz
Approved on 11 January 2013
Internal NMI service identifier: NIMT/14010 - 11101
 
Pressure sensitivity level. Measurement microphone type LS1P. Measurand unit, dB (reference: 1 V/Pa)
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB (reference: 1 V/Pa): 0.07
IEC 61094-2:1992
Frequency: 10 kHz
Approved on 11 January 2013
Internal NMI service identifier: NIMT/14010 - 11101
 
Pressure sensitivity level. Measurement microphone type LS2P. Measurand unit, dB (reference: 1 V/Pa)
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB (reference: 1 V/Pa): 0.10
IEC 61094-2:1992
Frequency: 20 Hz
Approved on 11 January 2013
Internal NMI service identifier: NIMT/14010 - 11102
 
Pressure sensitivity level. Measurement microphone type LS2P. Measurand unit, dB (reference: 1 V/Pa)
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB (reference: 1 V/Pa): 0.06
IEC 61094-2:1992
Frequency: 31.5 Hz
Approved on 11 January 2013
Internal NMI service identifier: NIMT/14010 - 11102
 
Pressure sensitivity level. Measurement microphone type LS2P. Measurand unit, dB (reference: 1 V/Pa)
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB (reference: 1 V/Pa): 0.04
IEC 61094-2:1992
Frequency: 63 Hz to 12.5 kHz
Approved on 11 January 2013
Internal NMI service identifier: NIMT/14010 - 11102
 
Pressure sensitivity level. Measurement microphone type LS2P. Measurand unit, dB (reference: 1 V/Pa)
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB (reference: 1 V/Pa): 0.05
IEC 61094-2:1992
Frequency: 16 kHz
Approved on 11 January 2013
Internal NMI service identifier: NIMT/14010 - 11102
 
Pressure sensitivity level. Measurement microphone type LS2P. Measurand unit, dB (reference: 1 V/Pa)
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB (reference: 1 V/Pa): 0.08
IEC 61094-2:1992
Frequency: 20 kHz
Approved on 11 January 2013
Internal NMI service identifier: NIMT/14010 - 11102
 
DC voltage sources: low value ranges (equal or below 10 V). DC voltage source, multifunction calibrator, voltage V, 0 V to 10 V
Absolute expanded uncertainty (k = 2, level of confidence 95%) in µV: (0.5 v + 2), v in V
Compare against DC standard
Internal NMI service identifier: NIMT/7
 
DC voltage sources: intermediate values (higher than 10 V to 1100 V). DC voltage source, multifunction calibrator, voltage V, > 10 V to 100 V
Absolute expanded uncertainty (k = 2, level of confidence 95%) in µV: (0.5 v + 15), v in V
Compare against DC standard
Internal NMI service identifier: NIMT/8
 
DC voltage sources: intermediate values (higher than 10 V to 1100 V). DC voltage source, multifunction calibrator, voltage V, > 100 V to 1000 V
Absolute expanded uncertainty (k = 2, level of confidence 95%) in µV: (0.8 v + 200), v in V
Compare against DC standard
Internal NMI service identifier: NIMT/9
 
AC-DC voltage transfer: AC-DC transfer difference at higher voltages (higher than 5 V). Thermal converters, AC-DC transfer standard, 7 V to 10 V
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 13
AC-DC voltage transfer
Voltage: 7 V, 10 V
Frequency: 50 kHz to 70 kHz
Internal NMI service identifier: NIMT/105
 
AC voltage up to 1000 V: sources. Multifunction calibrator, 300 mV to 700 mV
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 23
AC-DC voltage transfer
Voltage: 300 mV to 500 mV, 600 mV, 700 mV
Frequency: 30 Hz to 300 Hz
Internal NMI service identifier: NIMT/160
 
AC voltage up to 1000 V: meters. AC voltmeter, multimeter, 100 mV to 200 mV
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 38
AC-DC voltage transfer
Voltage: 100 mV, 200 mV
Frequency: 10 Hz to 20 Hz
Internal NMI service identifier: NIMT/216
 
AC voltage up to 1000 V: meters. AC voltmeter, multimeter, 100 mV to 200 mV
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 25
AC-DC voltage transfer
Voltage: 100 mV, 200 mV
Frequency: 30 Hz to 300 Hz
Internal NMI service identifier: NIMT/217
 
AC voltage up to 1000 V: meters. AC voltmeter, multimeter, 100 mV to 200 mV
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 25
AC-DC voltage transfer
Voltage: 100 mV, 200 mV
Frequency: 400 Hz to 30 kHz
Internal NMI service identifier: NIMT/218
 
AC voltage up to 1000 V: meters. AC voltmeter, multimeter, 100 mV to 200 mV
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 25
AC-DC voltage transfer
Voltage: 100 mV, 200 mV
Frequency: 50 kHz to 70 kHz
Internal NMI service identifier: NIMT/219
 
AC voltage up to 1000 V: meters. AC voltmeter, multimeter, 100 mV to 200 mV
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 41
AC-DC voltage transfer
Voltage: 100 mV, 200 mV
Frequency: 100 kHz
Internal NMI service identifier: NIMT/220
 
AC voltage up to 1000 V: meters. AC voltmeter, multimeter, 100 mV to 200 mV
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 240
AC-DC voltage transfer
Voltage: 100 mV, 200 mV
Frequency: 200 kHz to 500 kHz
Internal NMI service identifier: NIMT/221