Refine your search |
|
CMC AREA |
|
|
|
|
PHYSICS |
|
|
|
|
|
CHEMICAL MATERIAL |
|
|
|
|
GEOGRAPHIC LOCATION |
|
|
|
|
|
 |
Result of the search |
|
|
Your query 'cenam' produced 249 results
|
New search
|
| | | Mexico, CENAM (Centro Nacional de Metrologia) | |  | Complete CMCs in Length for Mexico (.PDF file)
|  | |
| Laser radiations. Stabilized laser of the mise en pratique: absolute frequency, 474 THz | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in kHz: 24 | | Optical beat frequency | | Internal NMI service identifier: CENAM/1 |
| | | Laser radiations. Other stabilized laser: vacuum wavelength , 633 nm | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in nm: 2E-09  | | Optical beat frequency | | Internal NMI service identifier: CENAM/2 |
| | | Pressure sensitivity level. Measurement microphone type LS2. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.2 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 20 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS2. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.11 | | Substitution IEC 1094-5 | | Frequency: 250 Hz | | Internal NMI service identifier: CENAM/510-F012-002 |
| | | Sound pressure level. Pistonphone, single frequency 250 Hz, 70 dB (reference: 20 µPa) to 130 dB (reference: 20 µPa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.09 | | Calibrated measurement microphone | | Microphone type: LS1P | | Internal NMI service identifier: CENAM/510-F012-003 |
| | | Sound pressure level. Pistonphone or sound calibrator, single frequency 125 Hz to 1 kHz, 70 dB (reference: 20 µPa) to 130 dB (reference: 20 µPa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.11 | | Calibrated measurement microphone | | Microphone type: LS2P | | Internal NMI service identifier: CENAM/510-F012-003 |
| | | Sound pressure level. Sound calibrator, single frequency 125 to 1 kHz, 70 dB (reference: 20 µPa) to 130 dB (reference: 20 µPa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.16 | | Calibrated measurement microphone | | Microphone type: WS2P | | Internal NMI service identifier: CENAM/510-F015-003 |
| | | Sound pressure level. Sound calibrator, multi-frequency, 70 dB (reference: 20 µPa) to 115 dB (reference: 20 µPa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.16 | | Calibrated measurement microphone | | Frequency: 125 Hz to 1 kHz | | Microphone type: LS2P | | Internal NMI service identifier: CENAM/510-F012-008 |
| | | Sound pressure level. Sound calibrator, multi-frequency, 70 dB (reference: 20 µPa) to 115 dB (reference: 20 µPa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.21 | | Calibrated measurement microphone | | Frequency: 31.5 Hz to 8 kHz | | Microphone type: LS2P | | Internal NMI service identifier: CENAM/510-F012-008 |
| | | Sound pressure response level. Sound level meters. Measurand unit, dB (reference: 20 µPa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.3 | | Comparison IEC651 and IEC804 | | Frequency: 250 Hz to 1 kHz | | Internal NMI service identifier: CENAM/510-F015-001 |
| | | Air-conduction response level. Audiometers. Measurand unit, dB HL (reference: threshold of hearing) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.38 | | IEC60645-1 | | Frequency: 125 Hz to 8 kHz | | Internal NMI service identifier: CENAM/510-F016-001 |
| | | Bone-conduction response level. Audiometers. Measurand unit, dB HL (reference: threshold of hearing) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.4 | | IEC60645-1 | | Frequency: 125 Hz to 4 kHz | | Internal NMI service identifier: CENAM/510-F016-001 |
| | | Force response level. Mechanical coupler system. Measurand unit, dB (reference: reference equivalent threshold force) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.3 | | Comparison IEC 60373 | | Frequency: 125 Hz to 4 kHz | | Internal NMI service identifier: CENAM/510-F012-011 |
| | | Mechanical impedance. Mechanical coupler system. Measurand unit, dB (reference: 1 N/(m/s) ) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.2 | | Comparison IEC 60373 | | Frequency: 125 Hz to 4 kHz | | Internal NMI service identifier: CENAM/510-F012-011 |
| | | Charge sensitivity (phase shift). Accelerometer, 0 ° to 360 ° | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in °: 2 | | Laser interferometry ISO 16063-11 | | Frequency: 50 Hz to 5 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F011-001 |
| | | Voltage sensitivity (phase shift). Acceleration measuring chain, 0 ° to 360 ° | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in °: 2 | | Laser interferometry ISO 16063-11 | | Frequency: 50 Hz to 5 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F011-001 |
| | | Pressure sensitivity level. Measurement microphone type LS1. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.04 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 31.5 Hz to 250 Hz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS1. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.03 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 500 Hz to 3.1 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS1. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.05 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 4 kHz to 5 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS1. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.06 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 6.3 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
|