Refine your search |
|
CMC AREA |
|
|
|
|
PHYSICS |
|
|
|
|
CHEMICAL MATERIAL |
|
|
|
|
CHEMICAL ANALYTE |
|
|
|
|
GEOGRAPHIC LOCATION |
|
|
|
|
|
 |
Result of the search |
|
| | | Mexico, CENAM (Centro Nacional de Metrologia) | |  | Complete CMCs in Acoustics, Ultrasound, Vibration for Mexico (.PDF file)
|  | |
| Voltage sensitivity (phase shift). Acceleration measuring chain, 0 ° to 360 ° | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in °: 2 | | Laser interferometry ISO 16063-11 | | Frequency: 50 Hz to 5 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F011-001 |
| | | Charge sensitivity (magnitude). Accelerometer. Measurand unit, C/(m/s2) | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 1 | | Reciprocity ASA S2.2-1959 ISO-16063-1 | | Frequency: 100 Hz to 4 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F010-001 |
| | | Charge sensitivity (magnitude). Accelerometer. Measurand unit, C/(m/s2) | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.5 | | Laser interferometry ISO 16063-11 | | Frequency: 50 Hz to 1 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F011-001 |
| | | Charge sensitivity (magnitude). Accelerometer. Measurand unit, C/(m/s2) | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.8 | | Laser interferometry ISO 16063-11 | | Frequency: 1 kHz to 5 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F011-001 |
| | | Charge sensitivity (magnitude). Accelerometer. Measurand unit, C/(m/s2) | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 2 | | Laser interferometry ISO 16063-11 | | Frequency: 5 kHz to 8 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F011-001 |
| | | Charge sensitivity (magnitude). Accelerometer. Measurand unit, C/(m/s2) | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 1.2 | | Comparison, ISO 5347-3 | | Frequency: 50 Hz to 5 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F012-001 |
| | | Voltage sensitivity (magnitude). Acceleration measuring chain. Measurand unit, V/(m/s2) | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.5 | | Laser interferometry ISO 16063-11 | | Frequency: 50 Hz to 1 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F011-001 |
| | | Voltage sensitivity (magnitude). Acceleration measuring chain. Measurand unit, V/(m/s2) | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 1.2 | | Comparison, ISO 5347-3 | | Frequency: 50 Hz to 5 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F012-001 |
| | | Voltage sensitivity (magnitude). Acceleration measuring chain. Measurand unit, V/(m/s2) | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 1.7 | | Comparison, ISO 5347-3 | | Frequency: 50 Hz to 5 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F012-015 |
| | | Acceleration. Acceleration measuring instrument, 10 m/s2 to 100 m/s2 | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.5 | | Laser interferometry ISO 16063-11 | | Frequency: 50 Hz to 1 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F011-001 |
| | | Acceleration. Acceleration measuring instrument, 10 m/s2 to 100 m/s2 | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.8 | | Laser interferometry ISO 16063-11 | | Frequency: 1 kHz to 5 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F011-001 |
| | | Acceleration. Acceleration measuring instrument, 10 m/s2 to 50 m/s2 | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 1.2 | | Comparison, ISO 5347-3 | | Frequency: 50 Hz to 5 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F019-001 |
| | | Acceleration. Acceleration measuring instrument, 10 m/s2 to 50 m/s2 | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 2 | | Comparison, ISO 5347-3 | | Frequency: 50 Hz to 1 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-G'03-002 |
| | | Acceleration. Calibrators, 10 m/s2 to 100 m/s2 | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 1 | | Comparison, ISO 5347-3 | | Frequency: 80 Hz to 160 Hz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F012-009 |
| | | Acceleration. Calibrators, 10 m/s2 to 100 m/s2 | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 2 | | Comparison, ISO 5347-3 | | Frequency: 50 Hz to 5 kHz | | Sinusoidal excitation | | Internal NMI service identifier: CENAM/510-F012-009 |
|