Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (80)
PHYSICS
Impedance up to the MHz range
AC voltage, current, and power (5)
DC voltage, current, and resistance (43)
Mass, mass standards (24)
Temperature (8)
GEOGRAPHIC LOCATION
SIM (80)
Costa Rica (80)
Result of the search
Your query 'Costa Rica' produced 80 results New search
[<< Prev] 3  4

 
Costa Rica, ICE-LMVE (Instituto Costarricense de Electricidad, Laboratorio Metrológico de Variables Eléctricas)
Complete CMCs in Electricity and Magnetism for Costa Rica (.PDF file)
DC current, DC current sources, intermediate values. Multifunction calibrator, current generator, > 0.1 mA to 1 mA
Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 23 to 59
Direct comparison
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/31201
 
DC current, DC current sources, intermediate values. Multifunction calibrator, current generator, 1 mA to 10 mA
Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 23 to 59
Direct comparison
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/31202
 
DC current, DC current sources, intermediate values. Multifunction calibrator, current generator, 10 mA to 100 mA
Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 35 to 70
Direct comparison
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/31203
 
DC current, DC current sources, intermediate values. Multifunction calibrator, current generator, 0.1 A to 1 A
Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 97 to 167
Direct comparison
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/31204
 
DC current, DC current sources, intermediate values. Multifunction calibrator, current generator, 1 A to 10 A
Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 235 to 418
Direct comparison
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/31205
 
DC current, DC current meters, low values. Multimeter, 0.01 mA to 0.1 mA
Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 102 to 656
Direct comparison
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/32101
 
DC current, DC current meters, intermediate values. Multimeter, > 0.1 mA to 1 mA
Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 43 to 107
Direct comparison
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/32201
 
DC current, DC current meters, intermediate values. Multimeter, 1 mA to 10 mA
Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 40 to 77
Direct comparison
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/32202
 
DC current, DC current meters, intermediate values. Multimeter, 10 mA to 100 mA
Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 53 to 117
Direct comparison
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/32203
 
DC current, DC current meters, intermediate values. Multimeter, 0.1 A to 1 A
Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 94 to 204
Direct comparison
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/32204
 
DC current, DC current meters, intermediate values. Multimeter, 1 A to 11 A
Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 404 to 840
Direct comparison
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/32205
 
AC voltage (up to the MHz range), AC voltage up to 1100 V, sources. Multifunction calibrator, source, oscilloscope calibrator, 0.01 V to 1000 V
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 72 to 31213
Direct comparison
Frequency: 10 Hz to 500 kHz
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/52101
Uncertainty table
 
AC voltage (up to the MHz range), AC voltage up to 1100 V, meters. AC voltmeter, multimeter, multifunction calibrator, 0.01 V to 1000 V
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 50 to 10667
Direct comparison
Frequency: 10 Hz to 1 MHz
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/52201
Uncertainty table
 
AC current, AC current up to 100 A, sources. Multifunction calibrator, source, 0.0001 A to 10 A
Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 392 to 8803
Direct comparison
Frequency: 10 Hz to 100 kHz
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/62101
Uncertainty table
 
AC current, AC current up to 100 A, meters. AC ammeter, multimeter, multifunction calibrator, 0.0001 A to 11 A
Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 113 to 5849
Direct comparison
Frequency: 10 Hz to 10 kHz
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/62201
Uncertainty table
 
AC power, AC power and energy, single phase (frequencies below or equal to 400 Hz). Watthour meter, wattmeter, 15 W to 12000 W
Relative expanded uncertainty (k = 2, level of confidence 95%) in µW/VA: 289
Comparison
Frequency: 60 Hz
Applied voltage: 120 V and 240 V
Current: (0.25 A at 120 V and 0.3 A at 240 V) to 50 A
Power factor: 1, 0.5 lead, 0.5 lag
Approved on 23 March 2010
Internal NMI service identifier: ICE-LMVE/71101
 
Costa Rica, LACOMET (Laboratorio Costarricense de Metrologia)
Complete CMCs in Mass and related quantities for Costa Rica (.PDF file)
Conventional mass. Mass standards, 1 mg
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mg: 0.002
Direct comparison
Temperature: 19 ºC to 21 ºC
Relative humidity: 45 % to 60 %
Approved on 26 November 2008
Internal NMI service identifier: LACOMET/MF-MA-PR-01
 
Conventional mass. Mass standards, 2 mg
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mg: 0.002
Direct comparison
Temperature: 19 ºC to 21 ºC
Relative humidity: 45 % to 60 %
Approved on 26 November 2008
Internal NMI service identifier: LACOMET/MF-MA-PR-01
 
Conventional mass. Mass standards, 5 mg
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mg: 0.002
Direct comparison
Temperature: 19 ºC to 21 ºC
Relative humidity: 45 % to 60 %
Approved on 26 November 2008
Internal NMI service identifier: LACOMET/MF-MA-PR-01
 
Conventional mass. Mass standards, 10 mg
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mg: 0.003
Direct comparison
Temperature: 19 ºC to 21 ºC
Relative humidity: 45 % to 60 %
Approved on 26 November 2008
Internal NMI service identifier: LACOMET/MF-MA-PR-01