Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs Chemistry (298)
CMCs General Physics (276)
PHYSICS >
AC voltage, current, and power
Impedance up to the MHz range
Radio frequency measurements (52)
Mass, mass standards (28)
Sound in air (23)
Dimensional metrology (20)
Density (20)
Pressure (18)
Vibration (16)
Fluid flow (16)
Torque, Viscosity, Hardness and Gravity (15)
Force (15)
Temperature (15)
Properties of detectors and sources (8)
CHEMICAL MATERIAL >
sediment (20)
organic solution, multi-component (20)
HNO3 in aqueous solution (19)
soil (16)
particulates (16)
CHEMICAL ANALYTE >
zinc
lead (9)
cadmium (8)
chromium (5)
acenaphthene (5)
naphthalene (5)
GEOGRAPHIC LOCATION
SIM (574)
Mexico (574)
Result of the search
Your query 'cenam' produced 574 results New search
[<< Prev] 28  29

 
Mexico, CENAM (Centro Nacional de Metrologia)
Complete CMCs in Electricity and Magnetism for Mexico (.PDF file)
Transmission coefficient: on coaxial magnitude. 2-port coaxial device, 20 dB to 50 dB
Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in dB: 0.036 to 0.081
Vector automatic network analyzer
Frequency: 1 GHz to 2 GHz
Connector: Type N
Approved on 09 October 2012
Internal NMI service identifier: CENAM/410-E002-(001, 002, 003, 004, 007)-CC
Uncertainty table
 
Transmission coefficient: on coaxial magnitude. 2-port coaxial device, 20 dB to 50 dB
Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in dB: 0.036 to 0.152
Vector automatic network analyzer
Frequency: 2 GHz to 18 GHz
Connector: Type N
Approved on 09 October 2012
Internal NMI service identifier: CENAM/410-E002-(001, 002, 003, 004, 007)-CC
Uncertainty table
 
Transmission coefficient: on coaxial magnitude. 2-port coaxial device, 50 dB to 70 dB
Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in dB: 0.191 to 1.659
Vector automatic network analyzer
Frequency: 45 MHz to 1 GHz
Connector: Type N
Approved on 09 October 2012
Internal NMI service identifier: CENAM/410-E002-(001, 002, 003, 004, 007)-CC
Uncertainty table
 
Transmission coefficient: on coaxial magnitude. 2-port coaxial device, 50 dB to 70 dB
Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in dB: 0.080 to 0.330
Vector automatic network analyzer
Frequency: 1 GHz to 2 GHz
Connector: Type N
Approved on 09 October 2012
Internal NMI service identifier: CENAM/410-E002-(001, 002, 003, 004, 007)-CC
Uncertainty table
 
Transmission coefficient: on coaxial magnitude. 2-port coaxial device, 50 dB to 70 dB
Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in dB: 0.074 to 0.231
Vector automatic network analyzer
Frequency: 2 GHz to 18 GHz
Connector: Type N
Approved on 09 October 2012
Internal NMI service identifier: CENAM/410-E002-(001, 002, 003, 004, 007)-CC
Uncertainty table
 
Transmission coefficient: on coaxial phase. 2-port coaxial device, -180 º to 180 º
Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in º: 1.78 to 6.85
Vector automatic network analyzer
Frequency: 45 MHz to 18 GHz
Connector: Type N
Approved on 09 October 2012
Internal NMI service identifier: CENAM/410-E002-(001, 002, 003, 004, 007)-CC
Uncertainty table
 
Transmission coefficient: on coaxial magnitude. 2-port coaxial device, 0 dB to 20 dB
Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in dB: 0.023 to 0.066
Vector automatic network analyzer
Frequency: 45 GHz to 18 GHz
Connector: Type N
Approved on 09 October 2012
Uncertainty table
 
RF power. Precision source, 1 mW
Relative expanded uncertainty (k = 2, level of confidence 95.45%) in mW/W: 4.000
Thermistor mount with DC substitution bridge
Frequency: 50 MHz, 1 GHz
Approved on 09 October 2012
Internal NMI service identifier: CENAM/E018-001C-CC
 
RF voltage. RF voltage meter, 0.5 V to 1 V
Relative expanded uncertainty (k = 2, level of confidence 95.45%) in mV/V: 0.01 to 0.03
RF to DC transfer
Frequency: 10 kHz to 3 MHz
Approved on 09 October 2012
Uncertainty value depends on frequency
Uncertainty table
 
DC voltage sources: single values. Solid state voltage standard, 1 V
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.3
Ratio technique from 1.018 V to 1 V via a DVM
Room temperature: 23 °C
Ambient pressure: 81 kPa
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E011-003-01
 
DC voltage sources: single values. Solid state voltage standard, 1.018 V
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.3
Differential measurement using a redundant design
Room temperature: 23 °C
Ambient pressure: 81 kPa
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E011-003-02
 
DC voltage sources: single values. Solid state voltage standard, 10 V
Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.3
Ratio technique from 1.018 V to 10 V via a DVM and a resistive divider
Room temperature: 23 °C
Ambient pressure: 81 kPa
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E011-003-03
 
Transmittance, diffuse, spectral. Spectrally-neutral material, 0 to 1
Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.003
Spectrophotometer
Wavelength: 240 nm to 900 nm
Bandwidth: 1 nm to 5 nm
Specific measurement conditions: normal incidence
Best measurement capability is valid only using PTB-certified white reference standard; the uncertainty increases for NPL certified references.
Internal NMI service identifier: CENAM/520-D006-004
 
Reflectance, diffuse, spectral. Spectrally-neutral material, 0.003 to 1
Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.002 to 0.003, varies with measurand
Spectrophotometer
Wavelength range: 380 nm to 780 nm
Bandwidth: 1 nm to 5 nm
Specific measurement conditions: 8°/d, 8°/t
Best measurement capability is valid only using PTB-certified white reference standard; the uncertainty increases for NPL certified references.
Internal NMI service identifier: CENAM/520-D006-003
 
Colour, surface, x, y, Y. General material. x and y, 0.0001 to 1
Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.006
Spectrophotometer
Specific measurement conditions: 8°/d, 8°/t
Type of material: reflecting material
Internal NMI service identifier: CENAM/520-D006-003
 
Colour, surface, x, y, Y. General material Y, 0 to 100
Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.21
Spectrophotometer
Specific measurement conditions: 8°/d, 8°/t
Type of material: reflecting material
Internal NMI service identifier: CENAM/520-D006-003
 
Wavelength. Optical spectrum analyser, 600 nm to 1600 nm
Absolute expanded uncertainty (k = 2, level of confidence ~95%) in nm: 0.03
Approved on 21 September 2005
Internal NMI service identifier: CENAM/520-D003-008
 
Correlated color temperature. Tungsten lamp, 2000 K to 3200 K
Absolute expanded uncertainty (k = 2, level of confidence ~95%) in K: 50 to 100, varies with measurand
Spectroradiometer
Internal NMI service identifier: CENAM/520-D005-003
 
Color, emitted, x, y. General source, no range given
Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.006
Spectroradiometer
Bandwidth: 2.5 nm
Type of source: tungsten lamp
Internal NMI service identifier: CENAM/520-D005-003
 
Fibre optics. Length. Single-mode optical fibre, 1 km to 15 km
Absolute expanded uncertainty (k = 2, level of confidence ~95%) in m: 0.1
Phase shift method
Wavelength: 1310 nm, 1550 nm, 1625 nm
Standard used: vectorial network analyzer
Approved on 29 June 2010