Your query 'cenam' produced 574 results
|
New search
|
| | | Mexico, CENAM (Centro Nacional de Metrologia) | |  | Complete CMCs in Electricity and Magnetism for Mexico (.PDF file)
|  | |
| Transmission coefficient: on coaxial magnitude. 2-port coaxial device, 20 dB to 50 dB | | Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in dB: 0.036 to 0.081 | | Vector automatic network analyzer | | Frequency: 1 GHz to 2 GHz | | Connector: Type N | | Approved on 09 October 2012 | | Internal NMI service identifier: CENAM/410-E002-(001, 002, 003, 004, 007)-CC | | Uncertainty table |
| | | Transmission coefficient: on coaxial magnitude. 2-port coaxial device, 20 dB to 50 dB | | Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in dB: 0.036 to 0.152 | | Vector automatic network analyzer | | Frequency: 2 GHz to 18 GHz | | Connector: Type N | | Approved on 09 October 2012 | | Internal NMI service identifier: CENAM/410-E002-(001, 002, 003, 004, 007)-CC | | Uncertainty table |
| | | Transmission coefficient: on coaxial magnitude. 2-port coaxial device, 50 dB to 70 dB | | Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in dB: 0.191 to 1.659 | | Vector automatic network analyzer | | Frequency: 45 MHz to 1 GHz | | Connector: Type N | | Approved on 09 October 2012 | | Internal NMI service identifier: CENAM/410-E002-(001, 002, 003, 004, 007)-CC | | Uncertainty table |
| | | Transmission coefficient: on coaxial magnitude. 2-port coaxial device, 50 dB to 70 dB | | Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in dB: 0.080 to 0.330 | | Vector automatic network analyzer | | Frequency: 1 GHz to 2 GHz | | Connector: Type N | | Approved on 09 October 2012 | | Internal NMI service identifier: CENAM/410-E002-(001, 002, 003, 004, 007)-CC | | Uncertainty table |
| | | Transmission coefficient: on coaxial magnitude. 2-port coaxial device, 50 dB to 70 dB | | Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in dB: 0.074 to 0.231 | | Vector automatic network analyzer | | Frequency: 2 GHz to 18 GHz | | Connector: Type N | | Approved on 09 October 2012 | | Internal NMI service identifier: CENAM/410-E002-(001, 002, 003, 004, 007)-CC | | Uncertainty table |
| | | Transmission coefficient: on coaxial phase. 2-port coaxial device, -180 º to 180 º | | Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in º: 1.78 to 6.85 | | Vector automatic network analyzer | | Frequency: 45 MHz to 18 GHz | | Connector: Type N | | Approved on 09 October 2012 | | Internal NMI service identifier: CENAM/410-E002-(001, 002, 003, 004, 007)-CC | | Uncertainty table |
| | | Transmission coefficient: on coaxial magnitude. 2-port coaxial device, 0 dB to 20 dB | | Absolute expanded uncertainty (k = 2, level of confidence 95.45%) in dB: 0.023 to 0.066 | | Vector automatic network analyzer | | Frequency: 45 GHz to 18 GHz | | Connector: Type N | | Approved on 09 October 2012 | | Uncertainty table |
| | | RF power. Precision source, 1 mW | | Relative expanded uncertainty (k = 2, level of confidence 95.45%) in mW/W: 4.000 | | Thermistor mount with DC substitution bridge | | Frequency: 50 MHz, 1 GHz | | Approved on 09 October 2012 | | Internal NMI service identifier: CENAM/E018-001C-CC |
| | | RF voltage. RF voltage meter, 0.5 V to 1 V | | Relative expanded uncertainty (k = 2, level of confidence 95.45%) in mV/V: 0.01 to 0.03 | | RF to DC transfer | | Frequency: 10 kHz to 3 MHz | | Approved on 09 October 2012 Uncertainty value depends on frequency | | Uncertainty table |
| | | DC voltage sources: single values. Solid state voltage standard, 1 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.3 | | Ratio technique from 1.018 V to 1 V via a DVM | | Room temperature: 23 °C | | Ambient pressure: 81 kPa | | Approved on 28 April 2011 | | Internal NMI service identifier: CENAM/E011-003-01 |
| | | DC voltage sources: single values. Solid state voltage standard, 1.018 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.3 | | Differential measurement using a redundant design | | Room temperature: 23 °C | | Ambient pressure: 81 kPa | | Approved on 28 April 2011 | | Internal NMI service identifier: CENAM/E011-003-02 |
| | | DC voltage sources: single values. Solid state voltage standard, 10 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.3 | | Ratio technique from 1.018 V to 10 V via a DVM and a resistive divider | | Room temperature: 23 °C | | Ambient pressure: 81 kPa | | Approved on 28 April 2011 | | Internal NMI service identifier: CENAM/E011-003-03 |
| | | Transmittance, diffuse, spectral. Spectrally-neutral material, 0 to 1 | | Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.003 | | Spectrophotometer | | Wavelength: 240 nm to 900 nm | | Bandwidth: 1 nm to 5 nm | | Specific measurement conditions: normal incidence | | Best measurement capability is valid only using PTB-certified white reference standard; the uncertainty increases for NPL certified references. | | Internal NMI service identifier: CENAM/520-D006-004 |
| | | Reflectance, diffuse, spectral. Spectrally-neutral material, 0.003 to 1 | | Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.002 to 0.003, varies with measurand | | Spectrophotometer | | Wavelength range: 380 nm to 780 nm | | Bandwidth: 1 nm to 5 nm | | Specific measurement conditions: 8°/d, 8°/t | | Best measurement capability is valid only using PTB-certified white reference standard; the uncertainty increases for NPL certified references. | | Internal NMI service identifier: CENAM/520-D006-003 |
| | | Colour, surface, x, y, Y. General material. x and y, 0.0001 to 1 | | Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.006 | | Spectrophotometer | | Specific measurement conditions: 8°/d, 8°/t | | Type of material: reflecting material | | Internal NMI service identifier: CENAM/520-D006-003 |
| | | Colour, surface, x, y, Y. General material Y, 0 to 100 | | Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.21 | | Spectrophotometer | | Specific measurement conditions: 8°/d, 8°/t | | Type of material: reflecting material | | Internal NMI service identifier: CENAM/520-D006-003 |
| | | Wavelength. Optical spectrum analyser, 600 nm to 1600 nm | | Absolute expanded uncertainty (k = 2, level of confidence ~95%) in nm: 0.03 | | Approved on 21 September 2005 | | Internal NMI service identifier: CENAM/520-D003-008 |
| | | Correlated color temperature. Tungsten lamp, 2000 K to 3200 K | | Absolute expanded uncertainty (k = 2, level of confidence ~95%) in K: 50 to 100, varies with measurand | | Spectroradiometer | | Internal NMI service identifier: CENAM/520-D005-003 |
| | | Color, emitted, x, y. General source, no range given | | Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.006 | | Spectroradiometer | | Bandwidth: 2.5 nm | | Type of source: tungsten lamp | | Internal NMI service identifier: CENAM/520-D005-003 |
| | | Fibre optics. Length. Single-mode optical fibre, 1 km to 15 km | | Absolute expanded uncertainty (k = 2, level of confidence ~95%) in m: 0.1 | | Phase shift method | | Wavelength: 1310 nm, 1550 nm, 1625 nm | | Standard used: vectorial network analyzer | | Approved on 29 June 2010 |
|