Refine your search |
|
CMC AREA |
|
|
|
|
PHYSICS |
|
|
|
|
CHEMICAL ANALYTE |
|
|
|
|
GEOGRAPHIC LOCATION |
|
|
|
|
|
 |
Result of the search |
|
| | | Mexico, CENAM (Centro Nacional de Metrologia) | |  | Complete CMCs in Acoustics, Ultrasound, Vibration for Mexico (.PDF file)
|  | |
| Pressure sensitivity level. Measurement microphone type LS2. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.2 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 20 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS2. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.11 | | Substitution IEC 1094-5 | | Frequency: 250 Hz | | Internal NMI service identifier: CENAM/510-F012-002 |
| | | Sound pressure response level. Sound level meters. Measurand unit, dB (reference: 20 µPa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.3 | | Comparison IEC651 and IEC804 | | Frequency: 250 Hz to 1 kHz | | Internal NMI service identifier: CENAM/510-F015-001 |
| | | Air-conduction response level. Audiometers. Measurand unit, dB HL (reference: threshold of hearing) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.38 | | IEC60645-1 | | Frequency: 125 Hz to 8 kHz | | Internal NMI service identifier: CENAM/510-F016-001 |
| | | Bone-conduction response level. Audiometers. Measurand unit, dB HL (reference: threshold of hearing) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.4 | | IEC60645-1 | | Frequency: 125 Hz to 4 kHz | | Internal NMI service identifier: CENAM/510-F016-001 |
| | | Mechanical impedance. Mechanical coupler system. Measurand unit, dB (reference: 1 N/(m/s) ) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.2 | | Comparison IEC 60373 | | Frequency: 125 Hz to 4 kHz | | Internal NMI service identifier: CENAM/510-F012-011 |
| | | Pressure sensitivity level. Measurement microphone type LS1. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.04 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 31.5 Hz to 250 Hz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS1. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.03 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 500 Hz to 3.1 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS1. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.05 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 4 kHz to 5 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS1. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.06 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 6.3 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS1. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.1 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 8 kHz to 9 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS1. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.15 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 10 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS2. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.05 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 63 Hz to 3.15 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS2. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.06 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 4 kHz to 8 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS2. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.1 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 9 kHz to 10 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS2. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.12 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 12.5 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
| | | Pressure sensitivity level. Measurement microphone type LS2. Measurand unit, dB (reference: 1 V/Pa) | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.14 | | Pressure reciprocity IEC 1094-2:1992 | | Frequency: 14 kHz to 18 kHz | | Internal NMI service identifier: CENAM/510-F013-001 |
|