Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (284)
PHYSICS
Sound in air
AC voltage, current, and power (15)
DC voltage, current, and resistance (3)
Density (20)
Dimensional metrology (20)
Fibre optics (3)
Fluid flow (16)
Force (15)
Frequency (5)
Impedance up to the MHz range (16)
Laser frequencies (2)
Mass, mass standards (28)
Other DC and low frequency measurements (1)
Photometry (7)
Pressure (18)
Properties of detectors and sources (8)
Properties of materials (4)
Radio frequency measurements (52)
Sound in water (1)
Temperature (15)
Time scale difference (4)
Torque, Viscosity, Hardness and Gravity (15)
Vibration (16)
GEOGRAPHIC LOCATION
SIM (284)
Mexico (284)
Result of the search
Your query 'cenam' produced 284 results New search
[<< Prev] 3  4  5 [Next >>]

 
Mexico, CENAM (Centro Nacional de Metrologia)
Complete CMCs in Thermometry for Mexico (.PDF file)
Temperature. Tungsten strip filament vacuum lamps, 800 °C to 1700 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.4 to 1.2
Radiance temperature determination as a function of filament current
Wavelength: 650 nm
Filament current
Approved on 29 June 2004
Internal NMI service identifier: CENAM/420-E008-004
 
Temperature. Disappearing filament optical pyrometer, 800 °C to 2200 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 6 to 9
Comparison to lamp
Wavelength: 650 nm
Filament current
Approved on 29 June 2004
Internal NMI service identifier: CENAM/420-E008-005
 
Temperature. Radiation thermometers, 50 °C to 400 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 2.4
Comparison to variable temperature blackbody
Radiance temperature
Spectral wavelength
Furnace heated electrically
SPRT, type S thermocouple
Approved on 29 June 2004
Internal NMI service identifier: CENAM/420-E008-001
 
Temperature. Monochromatic radiation thermometers, 800 °C to 2200 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.7 to 2.3
Comparison to lamps
Wavelength: 650 nm, 910 nm
Filament current
Approved on 29 June 2004
Internal NMI service identifier: CENAM/420-E008-002
 
Temperature. Type E, J, K, N, and T thermocouple, -38 °C to 200 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.3
Comparison to SPRT, determining measuring errors
Bath temperature: ethylenglycol, silicone oil and water bath, SPRT
Pre-determined value of inhomogeneity included in the CMC entry
Approved on 12 July 2007
Internal NMI service identifier: CENAM/420-E013-001
 
Temperature. Type T thermocouple, 200 °C to 400 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.5
Comparison to standard TC, determining measuring errors
Furnace temperature: dry block furnace (300 °C to 1100 °C ), potasium heat pipe, 81/2 digit multimeter
Pre-determined value of inhomogeneity included in the CMC entry
Approved on 12 July 2007
Internal NMI service identifier: CENAM/420-E013-001
 
Temperature. Resistance thermometer, -80 °C to 10 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.012
Industrial type
Comparison in bath (ethanol): sheath dimensions, diameter 3 mm to 7 mm, length >= 20 cm
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
Internal NMI service identifier: CENAM/420-E014-010
 
Temperature. Resistance thermometer, 10 °C to 80 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.01
Industrial type
Comparison in bath (water): sheath dimensions, diameter 3 mm to 7 mm, length >= 15 cm
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
Internal NMI service identifier: CENAM/420-E014-011
 
Temperature. Resistance thermometer, 80 °C to 250 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.02
Industrial type
Comparison in bath (oil): sheath dimensions, diameter 3 mm to 7 mm, length >= 30 cm
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
Internal NMI service identifier: CENAM/420-E014-011
 
Temperature. Digital (resistance) thermometer, -80 °C to 10 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.012
Industrial type
Comparison in bath (ethanol): sheath dimensions, diameter 3 mm to 7 mm, length >= 20 cm
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
Internal NMI service identifier: CENAM/420-E014-013
 
Temperature. Digital (resistance) thermometer, 10 °C to 80 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.01
Industrial type
Comparison in bath (water): sheath dimensions, diameter 3 mm to 7 mm, length >= 15 cm
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
Internal NMI service identifier: CENAM/420-E014-013
 
Temperature. Digital (resistance) thermometer, 80 °C to 250 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.02
Industrial type
Comparison in bath (oil): sheath dimensions, diameter 3 mm to 7 mm, length >= 30 cm
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
Internal NMI service identifier: CENAM/420-E014-013
 
Temperature. Liquid-in-glass Mercury, -30 °C to 10 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.04
Comparison
Silicone oil and water bath, SPRT, CA precision thermometry bridge: graduation 0.1 °C, total immersion
Approved on 12 July 2007
Internal NMI service identifier: CENAM/420-E006-001
 
Temperature. Liquid-in-glass Mercury, 80 °C to 250 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.08
Comparison
Silicone oil and water bath, SPRT, CA precision thermometry bridge: graduation 0.2 °C, total immersion
Approved on 12 July 2007
Internal NMI service identifier: CENAM/420-E006-003
 
Time scale difference. Local clock, -1 s to +1 s
Absolute expanded uncertainty (k = 2, level of confidence 95%) in ns: 2
Local clock vs. UTC(CNM)
Averaging time: 1 hour
Approved on 25 February 2009
 
Time scale difference. Local clock, -1 s to +1 s
Absolute expanded uncertainty (k = 2, level of confidence 95%) in ns: 20
Local clock vs. UTC
Averaging time: 5 days
This service requires the publication of the Circular T of the BIPM corresponding to dates when the calibration services were performed in order to know the time difference between the clock under calibration and UTC
Approved on 25 February 2009
 
Time scale difference. Remote clock, -1 s to +1 s
Absolute expanded uncertainty (k = 2, level of confidence 95%) in ns: 15
Remote clock vs. UTC(CNM)
Averaging time: 1 day
GPS common-view comparison using multichannel GPS receiver
Approved on 25 February 2009
 
Time scale difference. Remote clock, -1 s to +1 s
Absolute expanded uncertainty (k = 2, level of confidence 95%) in ns: 25
Remote clock vs. UTC
Averaging time: 5 days
GPS common-view comparison using multichannel GPS receiver
This service requires the publication of the Circular T of the BIPM corresponding to dates when the calibration services were performed in order to know the time difference between the clock under calibration and UTC
Approved on 25 February 2009
 
Frequency. Standard frequency source, 5 MHz
Relative expanded uncertainty (k = 2, level of confidence 95%) in Hz/Hz: 1.0E-13
Local frequency standard
Averaging time: 2 days
Approved on 25 February 2009
 
Frequency. Standard frequency source, 10 MHz
Relative expanded uncertainty (k = 2, level of confidence 95%) in Hz/Hz: 1.0E-13
Local frequency standard
Averaging time: 2 days
Approved on 25 February 2009