Refine your search |
|
CMC AREA |
|
|
|
|
PHYSICS |
|
|
|
|
CHEMICAL MATERIAL |
|
|
|
|
GEOGRAPHIC LOCATION |
|
|
|
|
|
 |
Result of the search |
|
| | | Mexico, CENAM (Centro Nacional de Metrologia) | |  | Complete CMCs in Mass and related quantities for Mexico (.PDF file)
|  | |
| Absolute pressure gas medium. Low pressure gauges, 1 kPa to 120 kPa | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in Pa: 0.5 to 2.5 | | Direct comparison with standard | | Temperature: 20 °C | | Internal NMI service identifier: CENAM/720-F002-002 |
| | | Absolute pressure gas medium. Pressure measuring devices, 0.0001 Pa to 1 Pa | | Relative expanded uncertainty (k = 2, level of confidence 95%): 9.00E-03 | | Direct comparison with standard | | Temperature: 23 °C | | Internal NMI service identifier: CENAM/720-F006-001 |
| | | Absolute pressure gas medium. Pressure measuring devices, 1 Pa to 10 Pa | | Relative expanded uncertainty (k = 2, level of confidence 95%): 3.50E-03 | | Direct comparison with standard | | Temperature: 23 °C | | Internal NMI service identifier: CENAM/720-F006-001 |
| | | Absolute pressure gas medium. Pressure measuring devices, 10 Pa to 100 Pa | | Relative expanded uncertainty (k = 2, level of confidence 95%): 2.50E-03 | | Direct comparison with standard | | Temperature: 23 °C | | Internal NMI service identifier: CENAM/720-F006-001 |
| | | Absolute pressure gas medium. Pressure measuring devices, 100 Pa to 1 000 Pa | | Relative expanded uncertainty (k = 2, level of confidence 95%): 2.50E-03 | | Direct comparison with standard | | Temperature: 23 °C | | Internal NMI service identifier: CENAM/720-F006-001 |
| | | Differential pressure gas medium. Low pressure gauges, 5.00E-03 kPa to 150 kPa | | Relative expanded uncertainty (k = 2, level of confidence 95%) in kPa: (15E-06 pl) + (15E-06 ps) | | Direct comparison with standard | | Temperature: 20 °C | | Line pressure pl: plmax = 150 kPa | | Standard pressure ps: pdiff = ps - pl | | Internal NMI service identifier: CENAM/720-F002-005 |
| | | Absolute pressure gas medium. Pressure measuring devices, 10 kPa to 175 kPa | | Relative expanded uncertainty (k = 2, level of confidence 95%): 1.50E-05 | | Direct comparison with standard | | Temperature: 20 °C | | Internal NMI service identifier: CENAM/720-F002-001, 002, 003, 004 |
| | | Absolute pressure gas medium. Pressure measuring devices, 175 kPa to 350 kPa | | Relative expanded uncertainty (k = 2, level of confidence 95%): 1.80E-05 | | Direct comparison with standard | | Temperature: 20 °C | | Internal NMI service identifier: CENAM/720-F002-001, 002, 003, 004 |
| | | Absolute pressure gas medium. Pressure measuring devices, 0.35 MPa to 1.75 MPa | | Relative expanded uncertainty (k = 2, level of confidence 95%): 2.60E-05 | | Direct comparison with standard | | Temperature: 20 °C | | Internal NMI service identifier: CENAM/720-F002-001, 002, 003, 004 |
| | | Absolute pressure gas medium. Pressure measuring devices, 1.75 MPa to 7 MPa | | Relative expanded uncertainty (k = 2, level of confidence 95%): 3.80E-05 | | Direct comparison with standard | | Temperature: 20 °C | | Internal NMI service identifier: CENAM/720-F002-001, 002, 003, 004 |
| | | Gauge pressure liquid medium. Pressure measuring devices, 0.2 MPa to 100 MPa | | Relative expanded uncertainty (k = 2, level of confidence 95%): 4.50E-05 | | Direct comparison with standard | | Temperature: 20 °C | | Internal NMI service identifier: CENAM/720-F001A-001, 720-F001A-002 |
| | | Gauge pressure liquid medium. Pressure measuring devices, 100 MPa to 200 MPa | | Relative expanded uncertainty (k = 2, level of confidence 95%): 5.00E-05 | | Direct comparison with standard | | Temperature: 20 °C | | Internal NMI service identifier: CENAM/720-F001A-001, 720-F001A-002 |
| | | Gauge pressure liquid medium. Pressure measuring devices, 200 MPa to 500 MPa | | Relative expanded uncertainty (k = 2, level of confidence 95%): 2.00E-4 | | Direct comparison with standard | | Temperature: 20 °C | | Internal NMI service identifier: CENAM/720-F001A-001, 720-F001A-002 |
|