| | | Costa Rica, ICE-LMVE (Instituto Costarricense de Electricidad, Laboratorio Metrológico de Variables Eléctricas) | |  | Complete CMCs in Electricity and Magnetism for Costa Rica (.PDF file)
|  | |
DC resistance, DC resistance meters, intermediate values. Multimeter, 1.9 to 100000000  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 8.7 to 102 | | Direct comparison | | Resistance, fixed values: (1.9, 10, 19, 100, 190) , (1, 1.9, 10, 19, 100, 190) k , (1, 1.9, 10, 19, 100) M | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/22201 | | Uncertainty table |
| | | DC voltage, DC voltage sources, low value ranges. DC voltage source, multifunction calibrator, 0.01 V to 0.1 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 21 to 161 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/11201 |
| | | DC voltage, DC voltage sources, low value ranges. DC voltage source, multifunction calibrator, 0.1 V to 1 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 6.2 to 21 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/11202 |
| | | DC voltage, DC voltage sources, low value ranges. DC voltage source, multifunction calibrator, 1 V to 10 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 4.9 to 6.2 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/11203 |
| | | DC voltage, DC voltage sources, intermediate values. DC voltage source, multifunction calibrator, > 10 V to 100 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 4.9 to 6.4 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/11301 |
| | | DC voltage, DC voltage sources, intermediate values. DC voltage source, multifunction calibrator, 100 V to 1000 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 6.4 to 15.6 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/11302 |
| | | DC voltage, DC voltage meters, intermediate values. DC voltmeter, multimeter, > 0.001 V to 0.01 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 48 to 416 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/12201 |
| | | DC voltage, DC voltage meters, intermediate values. DC voltmeter, multimeter, 0.01 V to 0.1 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 12 to 48 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/12202 |
| | | DC voltage, DC voltage meters, intermediate values. DC voltmeter, multimeter, 0.1 V to 1 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 5.8 to 12 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/12203 |
| | | DC voltage, DC voltage meters, intermediate values. DC voltmeter, multimeter, 1 V to 10 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 3.8 to 5.8 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/12204 |
| | | DC voltage, DC voltage meters, intermediate values. DC voltmeter, multimeter, 10 V to 100 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 3.8 to 5.5 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/12205 |
| | | DC voltage, DC voltage meters, intermediate values. DC voltmeter, multimeter, 100 V to 1000 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 5.5 to 7.0 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/12206 |
| | DC resistance, DC resistance standards and sources, low values. Fixed resistor, resistance box, 0.1 to 1  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 53 to 402 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21101 |
| | DC resistance, DC resistance standards and sources, high values. Fixed resistor, resistance box, > 1 M to 10 M | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 49 to 119 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21301 |
| | DC resistance, DC resistance standards and sources, high values. Fixed resistor, resistance box, 10 M to 100 M | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 398 to 468 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21302 |
| | DC resistance, DC resistance standards and sources, high values. Fixed resistor, resistance box, 20 M to 200 M | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 173 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21303 |
| | DC resistance, DC resistance standards and sources, high values. Fixed resistor, resistance box, 0.2 G to 2 G | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 231 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21304 |
| | DC resistance, DC resistance standards and sources, high values. Fixed resistor, resistance box, 2 G to 20 G | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 693 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21305 |
| | DC resistance, DC resistance standards and sources, high values. Fixed resistor, resistance box, 20 G to 200 G | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 924 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21306 |
| | DC resistance, DC resistance meters, intermediate values. Teraohmmeter, 10 M | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 286 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/22202 |
|