Refine your search |
|
CMC AREA |
|
|
|
|
PHYSICS |
|
|
|
|
GEOGRAPHIC LOCATION |
|
|
|
|
|
 |
Result of the search |
|
|
Your query 'Costa Rica' produced 39 results
|
New search
|
| | | Costa Rica, ICE-LMVE (Instituto Costarricense de Electricidad, Laboratorio Metrológico de Variables Eléctricas) | |  | Complete CMCs in Electricity and Magnetism for Costa Rica (.PDF file)
|  | |
DC resistance, DC resistance meters, intermediate values. Multimeter, 1.9 to 100000000  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 8.7 to 102 | | Direct comparison | | Resistance, fixed values: (1.9, 10, 19, 100, 190) , (1, 1.9, 10, 19, 100, 190) k , (1, 1.9, 10, 19, 100) M | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/22201 | | Uncertainty table |
| | | DC voltage, DC voltage sources, intermediate values. DC voltage source, multifunction calibrator, > 10 V to 100 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 4.9 to 6.4 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/11301 |
| | | DC voltage, DC voltage sources, intermediate values. DC voltage source, multifunction calibrator, 100 V to 1000 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 6.4 to 15.6 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/11302 |
| | | DC voltage, DC voltage meters, intermediate values. DC voltmeter, multimeter, > 0.001 V to 0.01 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 48 to 416 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/12201 |
| | | DC voltage, DC voltage meters, intermediate values. DC voltmeter, multimeter, 0.01 V to 0.1 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 12 to 48 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/12202 |
| | | DC voltage, DC voltage meters, intermediate values. DC voltmeter, multimeter, 0.1 V to 1 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 5.8 to 12 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/12203 |
| | | DC voltage, DC voltage meters, intermediate values. DC voltmeter, multimeter, 1 V to 10 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 3.8 to 5.8 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/12204 |
| | | DC voltage, DC voltage meters, intermediate values. DC voltmeter, multimeter, 10 V to 100 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 3.8 to 5.5 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/12205 |
| | | DC voltage, DC voltage meters, intermediate values. DC voltmeter, multimeter, 100 V to 1000 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 5.5 to 7.0 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/12206 |
| | DC resistance, DC resistance standards and sources, low values. Fixed resistor, resistance box, 0.1 to 1  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 53 to 402 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21101 |
| | DC resistance, DC resistance standards and sources, intermediate values. Fixed resistor, resistance box, > 1 to 10  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 18 to 53 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21201 |
| | DC resistance, DC resistance standards and sources, intermediate values. Fixed resistor, resistance box, 10 to 100  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 16 to 50 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21202 |
| | DC resistance, DC resistance standards and sources, intermediate values. Fixed resistor, resistance box, 100 to 1000  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 10 to 14 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21203 |
| | DC resistance, DC resistance standards and sources, intermediate values. Fixed resistor, resistance box, 1 k to 10 k | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 10 to 14 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21204 |
| | DC resistance, DC resistance standards and sources, intermediate values. Fixed resistor, resistance box, 10 k to 100 k | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 10 to 14 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21205 |
| | DC resistance, DC resistance standards and sources, intermediate values. Fixed resistor, resistance box, 100 k to 1000 k | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 16 to 30 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21206 |
| | DC resistance, DC resistance standards and sources, high values. Fixed resistor, resistance box, > 1 M to 10 M | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 49 to 119 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21301 |
| | DC resistance, DC resistance standards and sources, high values. Fixed resistor, resistance box, 10 M to 100 M | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 398 to 468 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21302 |
| | DC resistance, DC resistance standards and sources, high values. Fixed resistor, resistance box, 20 M to 200 M | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 173 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21303 |
| | DC resistance, DC resistance standards and sources, high values. Fixed resistor, resistance box, 0.2 G to 2 G | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 231 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21304 |
|