Refine your search |
|
CMC AREA |
|
|
|
|
PHYSICS |
|
DC voltage, current, and resistance (22) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
DC voltage, DC voltage sources, low value ranges |
DC voltage, DC voltage meters, intermediate values |
DC resistance, DC resistance standards and sources, high values |
DC resistance, DC resistance standards and sources, intermediate values |
|
|
|
GEOGRAPHIC LOCATION |
|
|
|
|
|
 |
Result of the search |
|
|
Your query 'Costa Rica' produced 22 results
|
New search
|
| | | Costa Rica, ICE-LMVE (Instituto Costarricense de Electricidad, Laboratorio Metrológico de Variables Eléctricas) | |  | Complete CMCs in Electricity and Magnetism for Costa Rica (.PDF file)
|  | |
DC resistance, DC resistance meters, intermediate values. Multimeter, 1.9 to 100000000  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 8.7 to 102 | | Direct comparison | | Resistance, fixed values: (1.9, 10, 19, 100, 190) , (1, 1.9, 10, 19, 100, 190) k , (1, 1.9, 10, 19, 100) M | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/22201 | | Uncertainty table |
| | | DC voltage, DC voltage sources, intermediate values. DC voltage source, multifunction calibrator, > 10 V to 100 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 4.9 to 6.4 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/11301 |
| | | DC voltage, DC voltage sources, intermediate values. DC voltage source, multifunction calibrator, 100 V to 1000 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 6.4 to 15.6 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/11302 |
| | DC resistance, DC resistance standards and sources, low values. Fixed resistor, resistance box, 0.1 to 1  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 53 to 402 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/21101 |
| | DC resistance, DC resistance meters, low values. Multimeter, 1  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 97 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/22101 |
| | DC resistance, DC resistance meters, intermediate values. Teraohmmeter, 10 M | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 286 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/22202 |
| | DC resistance, DC resistance meters, intermediate values. Teraohmmeter, 100 M | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 174 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/22203 |
| | DC resistance, DC resistance meters, intermediate values. Teraohmmeter, 1 G | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 235 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/22204 |
| | DC resistance, DC resistance meters, high values. Teraohmmeter, 10 G | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 694 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/22301 |
| | DC resistance, DC resistance meters, high values. Teraohmmeter, 100 G | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 929 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/22302 |
| | | DC current, DC current sources, low values. Multifunction calibrator, current generator, 0.01 mA to 0.1 mA | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 26 to 82 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/31101 |
| | | DC current, DC current sources, intermediate values. Multifunction calibrator, current generator, > 0.1 mA to 1 mA | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 23 to 59 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/31201 |
| | | DC current, DC current sources, intermediate values. Multifunction calibrator, current generator, 1 mA to 10 mA | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 23 to 59 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/31202 |
| | | DC current, DC current sources, intermediate values. Multifunction calibrator, current generator, 10 mA to 100 mA | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 35 to 70 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/31203 |
| | | DC current, DC current sources, intermediate values. Multifunction calibrator, current generator, 0.1 A to 1 A | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 97 to 167 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/31204 |
| | | DC current, DC current sources, intermediate values. Multifunction calibrator, current generator, 1 A to 10 A | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 235 to 418 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/31205 |
| | | DC current, DC current meters, low values. Multimeter, 0.01 mA to 0.1 mA | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 102 to 656 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/32101 |
| | | DC current, DC current meters, intermediate values. Multimeter, > 0.1 mA to 1 mA | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 43 to 107 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/32201 |
| | | DC current, DC current meters, intermediate values. Multimeter, 1 mA to 10 mA | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 40 to 77 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/32202 |
| | | DC current, DC current meters, intermediate values. Multimeter, 10 mA to 100 mA | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µA/A: 53 to 117 | | Direct comparison | | Approved on 23 March 2010 | | Internal NMI service identifier: ICE-LMVE/32203 |
|