Refine your search |
|
CMC AREA |
|
|
|
|
PHYSICS |
|
|
|
|
CHEMICAL MATERIAL |
|
|
|
|
|
CHEMICAL ANALYTE |
|
|
|
|
|
GEOGRAPHIC LOCATION |
|
|
|
|
|
 |
Result of the search |
|
|
Your query 'cenam' produced 552 results
|
New search
|
| | | Mexico, CENAM (Centro Nacional de Metrologia) | |  | Complete CMCs in Photometry and Radiometry for Mexico (.PDF file)
|  | |
| Colour, surface, x, y, Y. General material. x and y, 0.0001 to 1 | | Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.006 | | Spectrophotometer | | Specific measurement conditions: 8°/d, 8°/t | | Type of material: reflecting material | | Internal NMI service identifier: CENAM/520-D006-003 |
| | | Colour, surface, x, y, Y. General material Y, 0 to 100 | | Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.21 | | Spectrophotometer | | Specific measurement conditions: 8°/d, 8°/t | | Type of material: reflecting material | | Internal NMI service identifier: CENAM/520-D006-003 |
| | | Wavelength. Optical spectrum analyser, 600 nm to 1600 nm | | Absolute expanded uncertainty (k = 2, level of confidence ~95%) in nm: 0.03 | | Approved on 21 September 2005 | | Internal NMI service identifier: CENAM/520-D003-008 |
| | | Fibre optics. Length. Single-mode optical fibre, 1 km to 15 km | | Absolute expanded uncertainty (k = 2, level of confidence ~95%) in m: 0.1 | | Phase shift method | | Wavelength: 1310 nm, 1550 nm, 1625 nm | | Standard used: vectorial network analyzer | | Approved on 29 June 2010 |
| | | Luminous intensity. Tungsten lamp, 1 cd to 10000 cd | | Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1 | | Photometric bench and reference photometer | | Correlated colour temperature: 2856 K | | Internal NMI service identifier: CENAM/520-D008-001 |
| | | Illuminance. Tungsten lamp, 1 lx to 20000 lx | | Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1 | | Photometric bench and reference photometer | | Correlated colour temperature: 2856 K |
| | | Illuminance responsivity, tungsten source. Illuminance meter. Measurand unit, A/lx or V/lx or Reading/lx | | Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1 | | Photometric bench and reference photometer | | Illuminance: 50 lx to 3000 lx | | Correlated colour temperature: 2856 K | | Internal NMI service identifier: CENAM/520-D008-002 |
| | | Luminous flux. Tungsten lamp, 10 lm to 100 lm | | Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1.4 | | Integrating sphere | | Correlated colour temperature: 2600 K to 3100 K | | Internal NMI service identifier: CENAM/520-D008-005 |
| | | Luminous flux. Tungsten lamp, 100 lm to 5000 lm | | Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1.2 | | Integrating sphere | | Correlated colour temperature: 2600 K to 3100 K | | Internal NMI service identifier: CENAM/520-D008-005 |
| | | Luminance. Tungsten-based source, 10 cd/m2 to 40000 cd/m2 | | Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1 to 1.6 | | Integrating sphere source | | Correlated colour temperature: 2856 K | | Internal NMI service identifier: CENAM/520-D008-006 |
| | | Luminance responsivity. Luminance meter. Measurand unit, A/(cd/m2), V/(cd/m2) | | Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 2.5 | | Integrating sphere source | | Luminance: 10 cd/m2 to 40000 cd/m2 | | Type of source used: Integrating sphere source | | Internal NMI service identifier: CENAM/520-D008-003 |
| | | Responsivity. Fibre optic power meter. Measurand unit, A/W or V/W or Reading/W | | Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1.2 | | Comparison | | Wavelengths: 850 nm, 1300 nm, 1550 nm | | Bandwidth: 5 nm | | Power level: 18 µW to 100 µW | | Internal NMI service identifier: CENAM/520-D003-001, 002, 003, 004, 006 |
|