Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (287)
CMCs Chemistry (276)
PHYSICS >
Dimensional metrology
Radio frequency measurements (52)
Mass, mass standards (28)
Sound in air (23)
Density (20)
Pressure (18)
Vibration (16)
Fluid flow (16)
Impedance up to the MHz range (16)
Torque, Viscosity, Hardness and Gravity (15)
Force (15)
AC voltage, current, and power (15)
Temperature (15)
CHEMICAL MATERIAL >
organic solution, multi-component
HNO3 in aqueous solution (21)
sediment (18)
methanol (15)
soil (14)
particulates (14)
CHEMICAL ANALYTE >
acenaphthene
phenanthrene
lead (9)
cadmium (8)
zinc (7)
chromium (5)
nitrogen monoxide (5)
GEOGRAPHIC LOCATION
SIM (563)
Mexico (563)
Result of the search
Your query 'cenam' produced 563 results New search
[<< Prev] 28  29

 
Mexico, CENAM (Centro Nacional de Metrologia)
Complete CMCs in Electricity and Magnetism for Mexico (.PDF file)
Inductance: self inductance, low values. Fixed inductor, 100 µH
Relative expanded uncertainty (k = 2, level of confidence 95%) in mH/H: 0.60
Substitution using an LCR meter
Frequency: 1 kHz
Current: 30 mA
Approved on 28 April 2011
Internal NMI service identifier: CENAM/E009-010-01
 
Transmittance, diffuse, spectral. Spectrally-neutral material, 0 to 1
Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.003
Spectrophotometer
Wavelength: 240 nm to 900 nm
Bandwidth: 1 nm to 5 nm
Specific measurement conditions: normal incidence
Best measurement capability is valid only using PTB-certified white reference standard; the uncertainty increases for NPL certified references.
Internal NMI service identifier: CENAM/520-D006-004
 
Reflectance, diffuse, spectral. Spectrally-neutral material, 0.003 to 1
Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.002 to 0.003, varies with measurand
Spectrophotometer
Wavelength range: 380 nm to 780 nm
Bandwidth: 1 nm to 5 nm
Specific measurement conditions: 8°/d, 8°/t
Best measurement capability is valid only using PTB-certified white reference standard; the uncertainty increases for NPL certified references.
Internal NMI service identifier: CENAM/520-D006-003
 
Colour, surface, x, y, Y. General material. x and y, 0.0001 to 1
Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.006
Spectrophotometer
Specific measurement conditions: 8°/d, 8°/t
Type of material: reflecting material
Internal NMI service identifier: CENAM/520-D006-003
 
Colour, surface, x, y, Y. General material Y, 0 to 100
Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.21
Spectrophotometer
Specific measurement conditions: 8°/d, 8°/t
Type of material: reflecting material
Internal NMI service identifier: CENAM/520-D006-003
 
Wavelength. Optical spectrum analyser, 600 nm to 1600 nm
Absolute expanded uncertainty (k = 2, level of confidence ~95%) in nm: 0.03
Approved on 21 September 2005
Internal NMI service identifier: CENAM/520-D003-008
 
Correlated color temperature. Tungsten lamp, 2000 K to 3200 K
Absolute expanded uncertainty (k = 2, level of confidence ~95%) in K: 50 to 100, varies with measurand
Spectroradiometer
Internal NMI service identifier: CENAM/520-D005-003
 
Color, emitted, x, y. General source, no range given
Absolute expanded uncertainty (k = 2, level of confidence ~95%): 0.006
Spectroradiometer
Bandwidth: 2.5 nm
Type of source: tungsten lamp
Internal NMI service identifier: CENAM/520-D005-003
 
Fibre optics. Length. Single-mode optical fibre, 1 km to 15 km
Absolute expanded uncertainty (k = 2, level of confidence ~95%) in m: 0.1
Phase shift method
Wavelength: 1310 nm, 1550 nm, 1625 nm
Standard used: vectorial network analyzer
Approved on 29 June 2010
 
Luminous intensity. Tungsten lamp, 1 cd to 10000 cd
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1
Photometric bench and reference photometer
Correlated colour temperature: 2856 K
Internal NMI service identifier: CENAM/520-D008-001
 
Illuminance. Tungsten lamp, 1 lx to 20000 lx
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1
Photometric bench and reference photometer
Correlated colour temperature: 2856 K
 
Illuminance responsivity, tungsten source. Illuminance meter. Measurand unit, A/lx or V/lx or Reading/lx
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1
Photometric bench and reference photometer
Illuminance: 50 lx to 3000 lx
Correlated colour temperature: 2856 K
Internal NMI service identifier: CENAM/520-D008-002
 
Luminous flux. Tungsten lamp, 10 lm to 100 lm
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1.4
Integrating sphere
Correlated colour temperature: 2600 K to 3100 K
Internal NMI service identifier: CENAM/520-D008-005
 
Luminous flux. Tungsten lamp, 100 lm to 5000 lm
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1.2
Integrating sphere
Correlated colour temperature: 2600 K to 3100 K
Internal NMI service identifier: CENAM/520-D008-005
 
Luminance. Tungsten-based source, 10 cd/m2 to 40000 cd/m2
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1 to 1.6
Integrating sphere source
Correlated colour temperature: 2856 K
Internal NMI service identifier: CENAM/520-D008-006
 
Luminance responsivity. Luminance meter. Measurand unit, A/(cd/m2), V/(cd/m2)
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 2.5
Integrating sphere source
Luminance: 10 cd/m2 to 40000 cd/m2
Type of source used: Integrating sphere source
Internal NMI service identifier: CENAM/520-D008-003
 
Responsivity, laser, power. General detector. Measurand unit, A/W or V/W or Reading/W
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 0.03
Laser source
Wavelength: 633 nm
Power level: 0.1 mW
Type of detector: Si trap
Si, ECPR detectors can be measured, uncertainty evaluated on each case
Internal NMI service identifier: CENAM/520-D007-001
 
Responsivity. Fibre optic power meter. Measurand unit, A/W or V/W or Reading/W
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 1.2
Comparison
Wavelengths: 850 nm, 1300 nm, 1550 nm
Bandwidth: 5 nm
Power level: 18 µW to 100 µW
Internal NMI service identifier: CENAM/520-D003-001, 002, 003, 004, 006
 
Responsivity, spectral, power. Broad band detector. Measurand unit, A/W, V/W, Reading/W
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 0.8 to 0.62, varies with wavelength
Monochromator plus tungsten source
Wavelength range: 350 nm to 450 nm
Bandwidth: 4 nm
Power level: 0.4 µW to 4 µW
Internal NMI service identifier: CENAM/520-D007-001
 
Responsivity, spectral, power. Broad band detector. Measurand unit, A/W, V/W, Reading/W
Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 0.62 to 0.76, varies with wavelength
Monochromator plus tungsten source
Wavelength range: 450 nm to 770 nm
Bandwidth: 4 nm
Power level: 0.4 µW to 4 µW
Internal NMI service identifier: CENAM/520-D007-001