Refine your search |
|
CMC AREA |
|
|
|
|
PHYSICS |
|
|
|
|
CHEMICAL ANALYTE |
|
|
|
|
GEOGRAPHIC LOCATION |
|
|
|
|
|
 |
Result of the search |
|
| | | Thailand, NIMT (National Institute of Metrology (Thailand)) | |  | Complete CMCs in Electricity and Magnetism for Thailand (.PDF file)
|  | |
| DC voltage sources: low value ranges (equal or below 10 V). DC voltage source, multifunction calibrator, voltage V, 0 V to 10 V | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in µV: (0.5 v + 2), v in V | | Compare against DC standard | | Internal NMI service identifier: NIMT/7 |
| | | DC voltage sources: intermediate values (higher than 10 V to 1100 V). DC voltage source, multifunction calibrator, voltage V, > 10 V to 100 V | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in µV: (0.5 v + 15), v in V | | Compare against DC standard | | Internal NMI service identifier: NIMT/8 |
| | | DC voltage sources: intermediate values (higher than 10 V to 1100 V). DC voltage source, multifunction calibrator, voltage V, > 100 V to 1000 V | | Absolute expanded uncertainty (k = 2, level of confidence 95%) in µV: (0.8 v + 200), v in V | | Compare against DC standard | | Internal NMI service identifier: NIMT/9 |
| | | DC voltage sources: single values. Solid state voltage standard, 10 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.4 | | Compare against DC standard | | Internal NMI service identifier: NIMT/1 |
| | | DC voltage sources: single values. Solid state voltage standard, standard cell, 1.018 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.9 | | Compare against DC standard | | Internal NMI service identifier: NIMT/2 |
| | | DC voltage sources: low values (equal or below 10 V). DC voltage source, multifunction calibrator, 100 mV | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 5 | | Compare against DC standard | | Internal NMI service identifier: NIMT/3 |
| | | DC voltage sources: low values (equal or below 10 V). DC voltage source, multifunction calibrator, 1 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.6 | | Compare against DC standard | | Internal NMI service identifier: NIMT/4 |
| | | DC voltage sources: intermediate values (higher than 10 V to 1100 V). DC voltage source, multifunction calibrator, 100 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.5 | | Compare against DC standard | | Internal NMI service identifier: NIMT/5 |
| | | DC voltage sources: intermediate values (higher than 10 V to 1100 V). DC voltage source, multifunction calibrator, 1000 V | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µV/V: 0.7 | | Compare against DC standard | | Internal NMI service identifier: NIMT/6 |
| | DC resistance sources: low values (equal or below 1 ohm). Fixed resistor, 0.001  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 2.8 | | DCC bridge & range extender | | Current: 10 A to 100 A | | Oil bath temperature: 23 °C | | Internal NMI service identifier: NIMT/10 |
| | DC resistance sources: low values (equal or below 1 ohm). Fixed resistor, 0.01  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 2.3 | | DCC bridge & range extender | | Current: 1 A to 10 A | | Oil bath temperature: 23 °C | | Internal NMI service identifier: NIMT/11 |
| | DC resistance sources: low values (equal or below 1 ohm). Fixed resistor, 0.1  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 2.3 | | DCC bridge & range extender | | Current: 0.1 A to 1 A | | Oil bath temperature: 23 °C | | Internal NMI service identifier: NIMT/12 |
| | DC resistance sources: low values (equal or below 1 ohm). Fixed resistor, 1  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 1.5 | | DCC bridge | | Current: 100 A | | Oil bath temperature: 23 °C | | Internal NMI service identifier: NIMT/13 |
| | DC resistance sources: intermediate values (higher than 1 ohm to 1 Mohm). Fixed resistor, 10  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 1.6 | | DCC bridge | | Current: 30 mA | | Oil bath temperature: 23 °C | | Internal NMI service identifier: NIMT/14 |
| | DC resistance sources: intermediate values (higher than 1 ohm to 1 Mohm). Fixed resistor, 100  | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 2.2 | | DCC bridge | | Current: 10 mA | | Oil bath temperature: 23 °C | | Internal NMI service identifier: NIMT/15 |
| | DC resistance sources: intermediate values (higher than 1 ohm to 1 Mohm). Fixed resistor, 1 k | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 2.5 | | DCC bridge | | Current: 3 mA | | Oil bath temperature: 23 °C | | Internal NMI service identifier: NIMT/16 |
| | DC resistance sources: intermediate values (higher than 1 ohm to 1 Mohm). Fixed resistor, 10 k | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 2.4 | | DCC bridge | | Current: 1 mA | | Oil bath temperature: 23 °C | | Internal NMI service identifier: NIMT/17 |
| | DC resistance sources: intermediate values (higher than 1 ohm to 1 Mohm). Fixed resistor, 100 k | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 2.5 | | DCC bridge | | Current: 0.1 mA | | Oil bath temperature: 23 °C | | Internal NMI service identifier: NIMT/18 |
| | DC resistance sources: intermediate values (higher than 1 ohm to 1 Mohm). Fixed resistor, 1 M | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 18 | | Wheastone bridge | | Voltage: 1 V | | Internal NMI service identifier: NIMT/19 |
| | DC resistance sources: high values (higher than 1 Mohm). Fixed resistor, 10 M | | Relative expanded uncertainty (k = 2, level of confidence 95%) in µ / : 32 | | Wheastone bridge | | Voltage: 1 V | | Internal NMI service identifier: NIMT/20 |
|