Belarus, BelGIM (Belarussian State Institute for Metrology)
Complete CMCs in Thermometry for Belarus (.PDF file)
Temperature. Digital Thermometer IPRT, 20 °C to 80 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.01
Comparison
Temperature controlled bath: water
Approved on 06 September 2013
Internal NMI service identifier: COO-T-BelGIM-1049
 
Temperature. Digital Thermometer IPRT, 80 °C to 250 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.03
Comparison
Temperature controlled bath: oil
Approved on 06 September 2013
Internal NMI service identifier: COO-T-BelGIM-1050
 
Temperature. Digital Thermometer IPRT, 250 °C to 420 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.4
Comparison
Temperature controlled furnace: fluidised aluminia
Approved on 06 September 2013
Internal NMI service identifier: COO-T-BelGIM-1051
 
Temperature. Digital Thermometer Termocouple, 20 °C to 80 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.4
Comparison
Temperature controlled bath: water
Approved on 06 September 2013
Internal NMI service identifier: COO-T-BelGIM-1053
 
Temperature. Digital Thermometer Termocouple, 80 °C to 420 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.4
Comparison
Temperature controlled bath: oil
Approved on 06 September 2013
Internal NMI service identifier: COO-T-BelGIM-1054
 
Temperature. Digital Thermometer Termocouple, 250 °C to 420 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.6
Comparison
Temperature controlled furnace: fluidised aluminia
Approved on 06 September 2013
Internal NMI service identifier: COO-T-BelGIM-1055
 
Temperature. Liquid-in glass thermometer, 300 °C to 420 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.2
Comparison
Temperature controlled furnace: fluidised aluminia
Approved on 06 September 2013
Internal NMI service identifier: COO-T-BelGIM-1056
 
Temperature. Tungsten strip lamps, 962 °C to 2800 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 2.5 to 9
Compared with tungsten strip lamp
Effective wavelength: 0.47 µm to 0.7 µm
Approved on 19 November 2009
Internal NMI service identifier: COO-T-BelGIM-1036
 
Temperature. Tungsten strip lamps, 962 °C to 2800 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 4 to 17
Compared with tungsten strip lamp
Effective wavelength: 0.8 µm to 2 µm
Approved on 19 November 2009
Internal NMI service identifier: COO-T-BelGIM-1036
 
Temperature. Radiation temperature sensors with display unit, 962 °C to 2500 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 10 to 25
Compared with standard radiation temperature sensor
Wavelength: 2 µm to 18 µm
Approved on 19 November 2009
Internal NMI service identifier: COO-T-BelGIM-1037
 
Temperature. Radiation temperature sensors with display unit, 20 °C to 600 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.31 to 1.1
Compared with standard blackbody radiator
Wavelength: 2 µm to 18 µm
Approved on 19 November 2009
Internal NMI service identifier: COO-T-BelGIM-1038
 
Temperature. Tin cell, 231.928 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 1
Direct comparison
Temperature-controlled furnace: 3-zone
Approved on 19 November 2009
Internal NMI service identifier: COO-T-BelGIM-1004
 
Temperature. Zinc cell, 419.527 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 2
Direct comparison
Temperature-controlled furnace: 3-zone
Approved on 19 November 2009
Internal NMI service identifier: COO-T-BelGIM-1005
 
Temperature. Standard platinum resistance thermometer, 231.928 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 2
Tin fixed point
Temperature-controlled furnace: 3-zone
Approved on 19 November 2009
Internal NMI service identifier: COO-T-BelGIM-1010
 
Temperature. Standard platinum resistance thermometer, 419.527 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 3
Zinc fixed point
Temperature-controlled furnace: 3-zone
Approved on 19 November 2009
Internal NMI service identifier: COO-T-BelGIM-1011
 
Temperature. Industrial platinum resistance thermometer, -80 °C to 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in K: 0.5 to 0.05
Comparison method
Cryostat, baths, thermostats
Approved on 20 May 2010
Internal NMI service identifier: COO-T-BelGIM-1014
 
Temperature. Industrial platinum resistance thermometer, 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 5
Comparison method
Thermostat: ice bath
Approved on 20 May 2010
Internal NMI service identifier: COO-T-BelGIM-1015
 
Temperature. Industrial platinum resistance thermometer, 0 °C to 231 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in K: 0.02
Comparison method
Temperature-controlled bath, furnaces
Approved on 20 May 2010
Internal NMI service identifier: COO-T-BelGIM-1016
 
Temperature. Industrial platinum resistance thermometer, 156 °C to 231 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in K: 0.03
Comparison method
Temperature-controlled bath, furnaces
Approved on 20 May 2010
Internal NMI service identifier: COO-T-BelGIM-1017
 
Temperature. Industrial platinum resistance thermometer, 231 °C to 419 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in K: 0.05
Comparison method
Temperature-controlled bath, furnaces
Approved on 20 May 2010
Internal NMI service identifier: COO-T-BelGIM-1018