Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (11)
PHYSICS
Mass, mass standards
Temperature (11)
Temperature (11)
CHEMICAL ANALYTE
carbon monoxide
GEOGRAPHIC LOCATION
APMP (11)
Thailand (11)
Result of the search
Your query 'NIMT' produced 11 results New search

 
Thailand, NIMT (National Institute of Metrology (Thailand))
Complete CMCs in Thermometry for Thailand (.PDF file)
Temperature. Long stem SPRT, -38.8344 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 0.8
Triple point of Mercury
Temperature: seal cell and cryostat
Approved on 23 November 2009
Internal NMI service identifier: NIMT/11010-11320
 
Temperature. Long stem SPRT, 29.7646 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 0.8
Melting point of Gallium
Temperature: seal cell and thermostat
Approved on 23 November 2009
Internal NMI service identifier: NIMT/11010-11320
 
Temperature. Long stem SPRT, 231.9280 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 1.5
Freezing point of Tin
Temperature: seal cell and furnace
Approved on 23 November 2009
Internal NMI service identifier: NIMT/11010-11320
 
Temperature. Long stem SPRT, 419.527 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 2
Freezing point of Zinc
Temperature: seal cell and furnace
Approved on 23 November 2009
Internal NMI service identifier: NIMT/11010-11320
 
Temperature. IPRTs, -40 °C to 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 20
Comparison with SPRTs in ethanol bath
Approved on 16 March 2011
 
Temperature. IPRTs, 5 °C to 100 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 20
Comparison with SPRTs in water bath
Approved on 16 March 2011
 
Temperature. IPRTs, 100 °C to 250 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 20
Comparison with SPRTs in oil bath
Approved on 16 March 2011
 
Temperature. IPRTs, 250 °C to 420 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 30
Comparison with SPRTs in baths
Approved on 16 March 2011
 
Temperature. Pt-alloy thermocouples, 0 °C to 961.78 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.4 to 0.5
ITS-90 fixed point cells
Approved on 16 March 2011
 
Temperature. Pt-alloy thermocouples, 0 °C to 1100 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.5 to 1.5
Liquid bath + furnace
Approved on 16 March 2011
 
Temperature. Base metal thermocouple, 0 °C to 1100 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.5 to 2.5
Liquid bath + furnace
Approved on 16 March 2011