Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (8)
PHYSICS
Force
Fluid flow (8)
Volume of liquid (8)
CHEMICAL MATERIAL
organic solution, multi-component
CHEMICAL ANALYTE
chromium
GEOGRAPHIC LOCATION
SIM (8)
Mexico (8)
Result of the search
Your query 'cenam' produced 8 results New search

 
Mexico, CENAM (Centro Nacional de Metrologia)
Complete CMCs in Mass and related quantities for Mexico (.PDF file)
Volume of liquid. Volumetric test measure (micro-pipette), 0.01 cm3 to 1 cm3
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 1
Gravimetric determination
Temperature: 19 °C to 21 °C
Approved on 12 March 2007
Internal NMI service identifier: CENAM/FV9
 
Volume of liquid. Volumetric test measure (pipette, burette, flask), 1 cm3 to 10 cm3
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.05
Gravimetric determination
Temperature: 19 °C to 21 °C
Approved on 12 March 2007
Internal NMI service identifier: CENAM/FV10
 
Volume of liquid. Volumetric test measure (pipette, burette, flask), 10 cm3 to 2000 cm3
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.01
Gravimetric determination
Temperature: 19 °C to 21 °C
Approved on 12 March 2007
Internal NMI service identifier: CENAM/FV11
 
Volume of liquid. Volumetric test measure (pycnometer), 10 cm3 to 100 cm3
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.004
Gravimetric determination
Temperature: 19 °C to 21 °C
Approved on 12 March 2007
Internal NMI service identifier: CENAM/FV12
 
Volume of liquid. Volumetric test measure (graduated neck type), 2 dm3 to 50 dm3
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.015
Gravimetric determination
Temperature: 19 °C to 21 °C
Approved on 12 March 2007
Internal NMI service identifier: CENAM/FV13
 
Volume of liquid. Volumetric test measure (overflow type pipette), 5 dm3 to 50 dm3
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.005
Gravimetric determination
Temperature: 19 °C to 21 °C
Approved on 12 March 2007
Internal NMI service identifier: CENAM/FV14
 
Volume of liquid. Volumetric test measure (graduated neck type), 50 dm3 to 3000 dm3
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.02
Volumetric transfer
Temperature: 18 °C to 26 °C
Approved on 12 March 2007
Internal NMI service identifier: CENAM/FV15
 
Volume of liquid. Volume quantity meter, 50 dm3 to 3000 dm3
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.06
Turbine and positive displacement meter
Temperature: 18 °C to 26 °C
Kinematic viscosity: 0.5 mm2/s to 5 mm2/s
Flow: < 50 dm3/s
Approved on 12 March 2007
Internal NMI service identifier: CENAM/FV16